Note:
The instrument shown in these images is a CI-340 Photosynthesis Analyzer (CID Bio-
Science, Inc). The buffer tank system technique can be used to help stabilize the ambient intake
of any IRGA CO
2
sensor (images courtesy of ZealQuest, China).
CO2 PPM
CO2 PCT
Please follow the Standard Operating Procedure on page 29 and refer to Table 1 on page 27 to
calibrate this sensor.
The firmware supports a 3 point calibration for this particular sensor for added accuracy
between the 0-20% range. The sensor leaves our facility with the following 3 point span
calibration:
0.1% CO2
5% CO2
16%
If you run a span calibration for this sensor, it will replace the closest point from the 3 points
listed above.
O
2
Please follow the Standard Operating Procedure on page 29 and refer to Table 1 on page 27 to
calibrate this sensor.