Cleaning and Maintenance
Keep the lens of the F-750 clean and protected from dirt and scratches. If the lens is dirtied, unscrew
the cover plate and gently wipe the lens in a smooth straight motion with rubbing alcohol and a clean
nonabrasive cloth. Then, re-affix the cover plate. If the lens requires rinsing with water, ensure that the
rubber bumper (that protects water from leaking into the USB and SDcard slot) is snuggly installed. The
rubber bumper and the case of the F-50 can be wiped clean with a cloth.
Firmware Update Procedure
The firmware of the F-750 is updated by connecting the instrument to a computer using a USB cable and
Running the Firmware Update tool. The firmware update tool is available online at
http://www.felixinstruments.com/Software/F-750/FirmwareUpdateTool/
. For questions concerning
firmware updates, please contact
.
Warnings
Instrument will become warm with use, due to the lamp turning on.
Do not drop F-750 batteries on hard surface, such as cement floor.
Dropping the batteries in the 750 on a hard surface may cause them to crack the seal and rupture. The
electrolyte in the battery will slowly leak out; the electrolyte is toxic and may cause burns.