Fein Optic M46 User Manual Download Page 9

M46 Microscope Operation Instructions

8

Fein 

With the microscope powered on, turn the 

light adjustment knob (1) until the illumination 

is comfortable for observation. Turn the light 

adjustment knob clockwise to increase the light 

intensity and turn counter clockwise to decrease 

the light intensity.

Rest the light intensity by pressing the light 

intensity reset button (3). Adjust the light 

intensity by rotating the setting screw (2) with 

a compatible tool. Rotate clockwise to increase 

intensity and rotate counter clockwise to 

decrease intensity.

1

To select the light path, push the beam splitter (1) 

to the innermost position allowing the light to be 

directed to the eyepieces for viewing through the 

eyepieces. Pull the beam splitter to the outermost 

position allowing the light to be directed to the 

trinocular port for viewing with a microscope 

camera.

To adjust the focus, place a sample on the stage 

and position the 5x objective lens into the light 

path. Loosen the coarse adjusting limit knob (1) 

and adjust the left diopter adjustment ring to 

zero. Observe the right eyepiece with the right 

eye. Rotate the coarse focusing knob (2) until 

the sample appears in the field of view, then lock 

the coarse adjusting limit knob. Rotate the fine 

focusing knob (3) to bring the sample into crisper 

focus.

3

2

1

1

3

2

Loosen

Lock

Summary of Contents for M46

Page 1: ...Fein M46 Metallurgical Microscope Semiconductor Inspection User s Manual info feinoptic com...

Page 2: ...inclined trinocular viewing head erect image Reflected Light Frame High eyepoint eyepieces Brightfield Darkfield quintuple nosepiece Lamp House Power Switch X axis adjustment knob Y axis adjustment k...

Page 3: ...adjusting knob Aperture diaphragm adjusting knob Brightfield Darkfield selecting knob Aperture diaphragm centering screw Field diaphragm centering screw Illumination adjustment knob Clutch control kno...

Page 4: ...Connect the microscope head to the reflected illumination arm Insert the eyepieces into the eyetubes Attach the microscope stage to the microscope frame Screw the objective lenses into the nosepiece...

Page 5: ...lock piece 2 and remove the bulb 3 with clean a soft tissue Insert the bulb pins 4 into the jack on the bulb holder 5 The bulb should be vertical after assembly Loosen the bulb lock piece 2 the bulb...

Page 6: ...ing sure the clutch 4 is on the right hand side Move the top layer of the stage until the holes 3 appear Align holes 3 and 5 so they are superimposed Lock with screws Move the top layer of the stage i...

Page 7: ...e set screw 1 then position the microscope head into the dovetail on the illumination arm ensuring the eyepieces are facing forward Secure the head in place with the set screw 1 Low High Remove the ey...

Page 8: ...2 throughly Then screw down the nut 4 Insert one end of the power cord into the power socket 3 of the microscope Do not use strong force when the power cord is bent or twisted to avoid damage Only use...

Page 9: ...he light path push the beam splitter 1 to the innermost position allowing the light to be directed to the eyepieces for viewing through the eyepieces Pull the beam splitter to the outermost position a...

Page 10: ...re 5 diopters on the diopter adjustment ring with the value aligned with the scale of your eye s diopter The dot on the left side can also indicate the setting When using the diopter adjustable eyepie...

Page 11: ...the stage to move the sample to the center position aligned with the center of the objective If the position of the sample to be moved when the multi sample is observed is far from the center of the...

Page 12: ...ering screws 2 until the image of the field diaphragm is in the center of the field of view Gradually open the field diaphragm until the image of the field diaphragm is overlaid with the filed of view...

Page 13: ...nalyzer slot 4 insert the 360 rotating analyzer 3 or fixed analyzer 5 into the analyzer slot 4 until it is fully seated within the slot The polarizer and the analyzer are orthogonal when the 360 rotat...

Page 14: ...3 1 4 Install the c mount adapter to attach a microscope camera by loosening the lock screw 1 and remove the dust cap 2 Remove the two dust caps of the adapter 3 Insert the adapter into the trinocula...

Page 15: ...d of view Stains have accumulated on the specimen Clean the specimen Stains have accumulated on the lens Clean the lens Unclear Image The specimen is not vertical to the objective Adjust it The apertu...

Page 16: ...limit knob is locked Loosen the coarse limit knob The image moves obviously when touching the stage The stage is fastened incorrectly Fasten the stage correctly The bulb doesn t work No power supply...