Fein Optic M46 User Manual Download Page 10

M46 Microscope Operation Instructions

9

Fein 

If the handle is heavy when using the coarse 

focus or the specimen moves out of focus, it 

may be necessary to tighten the focus knob 

tension. This can be done by rotating the tension 

adjustment ring (1) according to the arrows.

1

After the image is clear in the right eyepiece, 

observe the specimen through the left eyepiece 

with the left eye. If the image is not in focus, 

rotate the diopter adjustment ring (1) until the 

image is clear. There are +/-5 diopters on the 

diopter adjustment ring with the value aligned 

with the scale of your eye’s diopter. The dot “.“ 

on the left side can also indicate the setting. 

When using the diopter adjustable eyepiece, first 

set the eyepiece diopter to zero. The eyepiece 

diopter can continue to be adjusted while 

operating the microscope to achieve a clear 

image.  

Loosen

Lock

1

Adjust the interpupillary distance by holding 

the base of the eyepiece on the left and right 

and pull them up or down to adjust to your 

sitting position. Hold the base of the eyepieces 

and rotate them around the axis to adjust the 

interpupillary distance until there is one field of 

view and comfortable for observation. 

Summary of Contents for M46

Page 1: ...Fein M46 Metallurgical Microscope Semiconductor Inspection User s Manual info feinoptic com...

Page 2: ...inclined trinocular viewing head erect image Reflected Light Frame High eyepoint eyepieces Brightfield Darkfield quintuple nosepiece Lamp House Power Switch X axis adjustment knob Y axis adjustment k...

Page 3: ...adjusting knob Aperture diaphragm adjusting knob Brightfield Darkfield selecting knob Aperture diaphragm centering screw Field diaphragm centering screw Illumination adjustment knob Clutch control kno...

Page 4: ...Connect the microscope head to the reflected illumination arm Insert the eyepieces into the eyetubes Attach the microscope stage to the microscope frame Screw the objective lenses into the nosepiece...

Page 5: ...lock piece 2 and remove the bulb 3 with clean a soft tissue Insert the bulb pins 4 into the jack on the bulb holder 5 The bulb should be vertical after assembly Loosen the bulb lock piece 2 the bulb...

Page 6: ...ing sure the clutch 4 is on the right hand side Move the top layer of the stage until the holes 3 appear Align holes 3 and 5 so they are superimposed Lock with screws Move the top layer of the stage i...

Page 7: ...e set screw 1 then position the microscope head into the dovetail on the illumination arm ensuring the eyepieces are facing forward Secure the head in place with the set screw 1 Low High Remove the ey...

Page 8: ...2 throughly Then screw down the nut 4 Insert one end of the power cord into the power socket 3 of the microscope Do not use strong force when the power cord is bent or twisted to avoid damage Only use...

Page 9: ...he light path push the beam splitter 1 to the innermost position allowing the light to be directed to the eyepieces for viewing through the eyepieces Pull the beam splitter to the outermost position a...

Page 10: ...re 5 diopters on the diopter adjustment ring with the value aligned with the scale of your eye s diopter The dot on the left side can also indicate the setting When using the diopter adjustable eyepie...

Page 11: ...the stage to move the sample to the center position aligned with the center of the objective If the position of the sample to be moved when the multi sample is observed is far from the center of the...

Page 12: ...ering screws 2 until the image of the field diaphragm is in the center of the field of view Gradually open the field diaphragm until the image of the field diaphragm is overlaid with the filed of view...

Page 13: ...nalyzer slot 4 insert the 360 rotating analyzer 3 or fixed analyzer 5 into the analyzer slot 4 until it is fully seated within the slot The polarizer and the analyzer are orthogonal when the 360 rotat...

Page 14: ...3 1 4 Install the c mount adapter to attach a microscope camera by loosening the lock screw 1 and remove the dust cap 2 Remove the two dust caps of the adapter 3 Insert the adapter into the trinocula...

Page 15: ...d of view Stains have accumulated on the specimen Clean the specimen Stains have accumulated on the lens Clean the lens Unclear Image The specimen is not vertical to the objective Adjust it The apertu...

Page 16: ...limit knob is locked Loosen the coarse limit knob The image moves obviously when touching the stage The stage is fastened incorrectly Fasten the stage correctly The bulb doesn t work No power supply...

Reviews: