Advice and declaration of the manufacturer: electromagnetic immunity
The SO
♥
CONNECT PID pump is designed for use in the electromagnetic environment specified below. The user of the SO
♥
CONNECT PID
pump must make sure that it is used in such an environment.
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment - Tips
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV (contact)
±8 kV (in the air)
± 8 kV contact
± 15kV air
Floors finished in wood, concrete or tiles. If the
floors are in synthetic material, the relative
humidity must be at least 30%.
Electrical
fast
transient/
burst
IEC 61000-4-4
± 2kV for supply
lines
± 2kV for supply
lines
Not applicable
Transient
surge
IEC 61000-4-5
± 1kV differential
mode
± 2kV differential
mode
Not applicable
Supply
lines
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Not applicable
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to
80 MHz outside
ISM bands
Not applicable
Radiated RF
IEC 61000-4-3
3V/m from 80MHz
to 2.5GHz
10V/m
Note (1)
Note (1): Portable and mobile RF communication devices may not be used at a distance from the components of the SO
♥
CONNECT PID pump
that is below the recommended distance, calculated on the basis of the equation applicable to the frequency of the emitter.
Recommended distance:
▪
d = (3/3.5)√P
▪
d = (3/3.5)√P from 80MHz to 800MHz
▪
d = (7/10)√P from 800 MHz to 2.5 GHz
where P is the maximum rated output power of the emitter, in watts (W) as indicated by the emitter manufacturer and d is the recommended
distance in metres.
The intensities of the fields from fixed RF emitters as determined by the electromagnetic reading on the site must be below the compliance level
in each reference range. Interference may occur close to equipment bearing the symbol:
Recommended separation distances from mobile/portable RF communication equipment
The SO
♥
CONNECT PID pump has been designed for use in an electromagnetic environment where the emitted RF interference is controlled.