Partie VIII
–
Caractéristiques techniques
68
Pump specifications
Pump dimensions
70.5 x 65 x 34.5 mm
Pump weight
146 g (battery included)
Battery
VARTA EZPack L 3.7V 1200 mAh 4.5 Wh
Battery life
2 infusions
(1)
Battery lifespan
> 500 cycles (>70% of initial capacity)
Disposable syringe
SO
♥
FILL 20 ml / SO
♥
FILL 30 ml / SO
♥
FILL 50 ml
Volumes that can be administered
Programmable from 1 ml to 50 ml in increments of 1 ml.
Infusion rates
Programmable from 0 to 60 ml/h, in increments of 0.1 ml/h.
Infusion rate precision
5%
Occlusion pressure
1.5 bar
Occlusion reporting time
See the appendix on the Time required for reporting an occlusion.
Post-occlusion bolus
0.5 ml
Memory
The infusion parameters are saved, even if the battery is removed.
Screen
TFT 2.4’’ RGB 320 x 240 pixels with resistive touchscreen
Motor
Stepping motor with magnetic encoder
Locking of reserved functions
Via password entry.
Protection rating
IPX2
Pump operating condition
0°C / +40°C
5% / 90% relative humidity without condensation
700 hPA / 1060 hPA
Pump storage condition
0°C / 50°C
5% / 90% relative humidity without condensation
700 hPa/1060 hPA
Acoustic range of sound alarm
44 dB(A) for high-priority alarms
38 dB(A) for low-priority alarms
Maximum volume infused in first fault condition
0.5% of the infusion rate/hour