Chapter 2: CAM2® Measure Devices Menu
N
OTE
: Seven-axis FaroArms have an additional position. Rotate the
handle.
CAUTION
: The probe
must
be well-seated in the hole when
digitizing all calibration points. Even one or two poorly digitized
points significantly affects the optimization process, which then has
an effect on the accuracy of the FaroArm.
5 Rotate the Probe to the next
position. Look at the dialog
box and move the handle
until you reach the start of
the next position.
• Repeat steps 1 through 4
and digitize 200 more
points in this position.
6 Rotate the Probe to the last
position. Look at the dialog
box and move the handle
until you reach the start of
the next position.
• Repeat steps 1 through 4
and digitize 200 more
points in this position.
Figure 2-6 Additional Position for Seven-axis FaroArms
08M46E00_FaroArmUSB.book Page 82 Tuesday, February 17, 2009 1:42 PM