It means the following:
A05
– The avg. is based on 5 measures
28,8
- The avg. moisture % of the
results.
If you don't want to include the results
just wait and the meter will automatically
switch off. Then just start again.
Before using the average calculation,
always check that the average memory
is empty. Delete if necessary.
4.1.2 Delete the average memory
Method 1
. When the average value is
displayed, press F and keep it pressed
until you see
0
and after that
A00
Method 2
. Press F
and keep it
depressed
then press
P
. When you see
A
on the display, release the
F
button. If
you now see an average value, you can
erase it by keeping the button
F
depressed until you see on the display
0
and after that
A00
.
Please note! It is a good practice to
erase the average memory after
completing a measuring session. This
prevents any old values affecting later
measurements.
The avg. memory blinks If it is full.
4.2 Adjusting the calibration
If you have had a sample tested in an
oven and the moisture is different to
what the meter reads. Then you can
adjust your meter match the oven test,
and this adjustment can remain in the