To change the measurement units (L/GAL/PT/QT meter only)

• 

Move the flashing display to Area 2 by pressing MENU, then press MENU again 

to choose measurement

• 

Press MENU for over three seconds to exit the setting mode

CAUTION 

• 

To be sure the proper amount of fluid is dispensed, always use 

the same measurement unit for a particular fluid. Units should be 

changed only by authorized employees.

• 

To verify the accuracy of a digital meter, use a clean, calibrated 

container. If using a single container, be sure to clean it after each 

dispense

• 

Have pump air pressure at the lowest possible setting for dispensing 

fluid

• 

Put the tip of the nozzle at the bottom of the calibrated container

• 

If the tip of the dispense valve does not reach the bottom of the 

calibrated container, use a length of plastic tubing over the tip of 

the nozzle to ensure liquid enters the container from the bottom

• 

Trigger the gun slowly so the fluid immediately covers the tip of the 

dispense valve

To activate the digital display

Press the RESET key pad to clear the meter before starting a new dispense cycle. 

This is the best way to activate the meter, because it also clears the quantity of the 

last dispense cycle.

The digital display can also be activated by pressing the MENU key pad or by run

-

ning fluid through the meter.

Calibrating automatically

• 

Press the RESET key for one second and area 1 shows .000 GAL

• 

Keep running fluid through the meter. The display will show a value as *.** 

GAL. Press the MENU key for three seconds and the meter will enter instrument 

calibration mode

• 

Press MENU to move from the selected number, then press RESET to choose a 

number. Enter the fluid volume and press MENU again for three seconds. 

Show current correction factor

• 

Press MENU and RESET at the same time to show the current correction factor. 

This number can be reset

Reset the resettlement total

• 

Press MENU for 10 seconds. The accumulated total will be reset to “0”

Summary of Contents for APX Industrial DCM-01

Page 1: ...Manual 800 247 1854 www FarmChem com DCM 01 Digital Manual Meter...

Page 2: ...e electrical and safety regulations INJECTION HAZARD Spray from the valve leaks or ruptured components can inject fluid into your body and cause serious injury including the need for amputation Fluid...

Page 3: ...for assistance in designing a system to suit your particular needs Pre Installation Procedure Relieve the pressure 1 Close the fluid shut off valve 2 Ground the hose and reel or console Do not use PTF...

Page 4: ...sening the 4 swivel nut then by turn ing the nozzle to your desired position and tightening the nut Existing Installation 1 Relieve the pressure 2 Loosen and disconnect the hose from the old dispense...

Page 5: ...essor Follow manufacturer s recommendations Fluid supply container Follow your local code To maintain grounding continuity when flushing or relieving pressure always hold a metal part of the valve fir...

Page 6: ...s the container from the bottom Trigger the gun slowly so the fluid immediately covers the tip of the dispense valve To activate the digital display Press the RESET key pad to clear the meter before s...

Page 7: ...ous injury including fluid injection never exceed the maximum working pressure of the valve you are using or of the low est rated component in your system DISPENSING PROCEDURE NOTE Before you begin ma...

Page 8: ...FarmChem www FarmChem com 8 EXPLODED AND PARTS LIST...

Page 9: ...zle MT10001 Manual tip MT10002 Auto tip MT10003 Semi auto tip Wearing parts Code Name Quantity 1 5 O ring 2 1 7 Washer flat 2 1 10 Seat 1 1 13 Filter 1 3 1 O ring 1 Code Name Quantity 3 5 Main circuit...

Page 10: ...ge or obstructions and replace if damaged 2 If the problem remains contact your distributor for repair or replacement Valve leaks O rings or valve seat are worn or damaged 1 Replace the o rings and or...

Page 11: ...800 247 1854 FarmChem 11...

Page 12: ......

Reviews: