9.
Slurry cups are tapered, with the large end at the top. This makes it easier to
press the cement plug out of the slurry cup. Press the cement plug out from
bottom to top. Take care to not bend the paddle shaft.
10.
Clean all surfaces and threads using a mild detergent. Be very careful not to
damage the threads or sealing surfaces of components as this may make
reassembly difficult.
11.
Flat diaphragms are generally thrown away after every test depending on
the maximum temperature reached during a test.
12.
Check the potentiometer for proper operation. Lightly re-grease threads.