Model 422 & Model 422CC Instruction Manual
D01187869
Revision A, December 2015
17
power adapter must be connected to the back of the instrument to prevent
instrument malfunction. See section 4 for installation information.
5.4
Software Main Menu
The main menu is starting point for the instrument. From here users may start new
tests, set up test parameters, setup instrument parameters, or view an old test. Also
located on the main menu is the reading for current temperature, consistency, and
pressure. On the right side of the screen, the version software that is currently
installed on the instrument is displayed. In the screenshot above, the version is
5.3.53.
•
The cooling button is located on the main screen and is operated by
selecting it to turn it on. Select again to turn off. The cooling button is
connected to a solenoid valve which is connected to the cooling input
located on the back of the instrument. Typically, water is used at house
pressure and temperature. However, circulating chiller can also be
connected. If cooling is turned on when start test is activated, the cooling
will remain on and heating will not ensue. Turn cooling off if starting a
normal test is desired.
•
The motor switch is located on the main screen and is operated by selecting
it to turn it on. Select again to turn off. The motor speed is set to turn at 150
rpm.
•
The fill button is located on the main screen. This button operates several
functions in one. As shown below, the fill button shows fill off, fill on, run
test, and drain cylinder. The fill off position neither allows oil or air into the
cylinder. The fill on position allows oil to flow from the reservoir into the