A-21
Electrical ratings
Rated supply voltage
Rated current input
Rated supply frequency
100–115 V
1.2 A (2.5 A max.)
50–60 Hz
220–240 V
1.1 A (2.3 A max.)
50–60 Hz
Outlet air temperature
Maximum = 38 °C (100 °F)
Noise level
Sound pressure level 28 ±1.5 dBA; average sound power level <35 dBA.
Water chamber volume
380 mL up to the maximum water-level line
Standards compliance
IEC 60601-1:2012; IEC 60601-1-2:2014; IEC 60601-1-2:2007; IEC 60601-1-11:2015;
ISO 80601-2-70:2015; ISO 5356-1:2004; ISO 17510-1:2007; ISO 8185:2007
FCC compliance
This device has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
device generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this device does cause harmful interference to
radio or television reception, which can be determined by turning the device off and
on, the user is encouraged to try to correct the interference by one or more of the
following measures:
• Reposition or relocate the receiving antenna.
• Increase the separation between the device and receiver.
• Connect the device into an outlet on a circuit different from that to which the
receiver is connected.
• Consult your healthcare provider or your Fisher & Paykel Healthcare
representative for help.
Data recording
The InfoUSB will store up to 5 years of summary efficacy data, 365 days of detailed
efficacy data, and 140 hours of high-resolution pressure, leak and flow data. Without
an InfoUSB, the device’s internal memory is capable of storing up to 1 year of summary
efficacy data, 30 days of detailed efficacy data, and 20 hours of high-resolution pressure,
leak, and flow data.
Service life
Device
5 years
Breathing tubes
12 months
Water chamber
12 months
Air filter
3 months
General
The patient is an intended operator.
8.4 CLASSIFICATIONS
Mode of operation
Continuous operation
Electric shock protection
Type BF
Ingress protection
IP22