Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3V
150 kHz to 80 MHz
6 V in ISM and
amateur radio
bands between 0,15
MHz and 80 MHz.
a
6 V in ISM and
amateur radio
bands between 0,15
MHz and 80 MHz.
a
80 MHz to 800 MHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
800 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2)
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2)
where p is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
b
should be less than the compliance
level in each frequency range.
c
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Proximity magnetic
fields
IEC 61000-4-39
30 kHz
,
CW
,
8A/m
(
This test is
applicable only to
ME EQUIPMENT
and ME SYSTEMS
intended for use in
the HOME
HEALTHCARE
ENVIRONMENT
)
134,2 kHz
,
Pulse
modulation 2,1
kHz
65 A/m
13,56 MHz Pulse
modulation 50 kHz
7,5 A/m
N/A
The FDBP-A10BL does not contain
magnetically sensitive components.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects and people.