Warning:
·
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
·
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this the FDBP-A10BL
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in
improper operation
·
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the FDBP-A10BL, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Attachment:
Guidance and manufacturer
´
s declaration
–
electromagnetic emission
–
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer
´
s declaration
–
electromagnetic emission
The FDBP-A10BL is intended for use in the electromagnetic environment specified below. The customer or the
user of FDBP-A10BL should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The FDBP-A10BL uses RF energy only for its internal function. There
for, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The FDBP-A10BL is suitable for use in all establishments,
including domestic establishments and those directly connected
to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
N/A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
N/A
Guidance and manufacturer's declaration
–
electromagnetic immunity
–
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer
´
s declaration
–
electromagnetic immunity
The FDBP-A10BL is intended for use in the electromagnetic environment specified below. The customer or the
user of the FDBP-A10BL should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -guidance