Electrostatic
discharge (ESD)
IEC 61000-4-2
±
8 kV contact
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV air
±
8 kV contact
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %. If ESD interfere with the operation
of equipment, counter measurements such
as wrist strap, grounding shall be considered.
Electrostatic
transient / burst
IEC 61000-4-4
±
2 kV for power
supply lines
N/A
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common
mode
N/A
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
(100 % dip in
UT) for 0,5 cycle
0 % U
T
(100 % dip in
UT) for 1 cycles
70 % U
T
(30 % dip in
UT) for 25/30cycles
0 % U
T
(100 % dip in
UT) for 250/300
cycles
N/A
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the FDBP-A10BL requires
continued operation during power mains
interruptions, it is recommended that the
FDBP-A10BL be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NOTE
:
U
T
is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer
´
s declaration
–
electromagnetic immunity
–
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer
´
s declaration
–
electromagnetic immunity
The FDBP-A10BL is intended for use in the electromagnetic environment specified below. The customer or the
user of the FDBP-A10BL should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the FDBP-A10BL, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance: