9.
Copy the lice
nse activation key into the top box “Please enter license key” and click the close
button.
Analyzer: How to connect to your device and operate the analyzer
Step 1: Connect the SMA cables for analyzer use
Note: the labels on the analyzer front panel are from the instrument perspective.
A.
Connect the TxDN from the device M-Tx lane 1. Connect the TxDP from the device M-Tx lane
1
. This is the “Up” direction.
B.
Connect the TxDN from the device M-Tx lane 0. Connect the TxDP from the device M-Tx lane
0. This
is the “Up” direction.
C.
Connect the RxDN from the device M-Rx lane 1. Connect the RxDP from the device M-Rx lane
1
. This is the “Down”
direction. This is used for the analyzer only, for the Exerciser these
connectors are not used.
D.
Connect the RxDN from the device M-Rx lane 0. Connect the RxDP from the device M-Rx lane
0
. This is the “Down” direction. This is used for the
analyzer only, for the Exerciser these
connectors are not used.
G.
Generates a UFS reference clock on 19.2 MHz.
H.
Generates a 1.8V signal to assert the UFS Hard Reset.
I.
Trigger in to the analyzer, 1.8V required.
J.
Trigger out from the analyzer, 1.8V is generated.
K.
Ground