Appendix 1 - Using the test port for the Dräger oil tube
As well as the Dräger oil Impactor the F8100 test port can also be used with the Dräger oil tube.
Test
Dräger ref.
Factair Part
No:
Sensitivity
Oil (chemical reagent tube)
6728371
TUB0003
0.1 mg/m
3
To prepare the tube Factair recommends the F2187 Dräger tube tip cutter.
If you have another Dräger tube, which you wish to use with the instrument, contact Factair for
technical advice on its suitability.
To complete a tube test, select a tube test
option as shown, press the button to proceed.
If you can determine with a high degree of
certainty the specific compressor lubricant
being used, then refer to the “test times” chart
provided on Factair’s website marked for the
F8100. If you have no knowledge of the type of
compressor lubricant being used in the
breathing-air system then Factair recommends
15 minutes as this covers the vast majority of
lubricants.
Once the instrument has completed its purge
sequence you will be prompted to insert the
tube into the test port and tighten firmly.
Note: Both ends of the tube should only be cut
immediately prior to the test. Ensure the arrow
printed on the tube is pointing away from the
instrument.
Press the right hand arrow to proceed.