"UPOC" Oxygen Concentrator 
 
 
Model: 
OC503-BL

OC503-BW

OC503-SR

OC503-TG

OC503-GY 

 

" UPOC " Oxygen Concentrator USER MANUAL 

Summary of Contents for OC503-BL

Page 1: ...UPOC Oxygen Concentrator Model OC503 BL OC503 BW OC503 SR OC503 TG OC503 GY UPOC Oxygen Concentrator USER MANUAL...

Page 2: ...G INSTRUCTIONS 11 INTRODUCTION 13 STRUCTURE 13 SCOPE OF APPLICATION 13 CONTRAINDICATION 13 ENVIRONMENT REQUIREMENTS 13 TRANSPORTATION AND STORAGE CONDITIONS 14 ADDITIONAL REMARKS 14 SPECIFICATIONS 16...

Page 3: ...hooting information repairment instructions etc In order for you to operate this machine effectively please read and understand the instructions in this manual carefully before using it Please note th...

Page 4: ...flame Improper use of the cord and plug may cause fire or other electric shocks and burns Please DO NOT use the machine with damaged AC power cord Before cleaning the dust of the Oxygen Concentrator...

Page 5: ...y viruses or bacteria between users Please use the provided humidifier bottle Do not replace it before consulting the supplier or doctor otherwise it may cause discomfort or difficulties with inhaling...

Page 6: ...vice s AC power cord with an extension cord Do not use other electrical appliances simultaneously on the same plug when you are operating the oxygen concentrator This machine is strictly prohibited in...

Page 7: ...rate and test or else it can cause damage to the machine NOTICE You must pay special attention to the information The purity will be up to 90 after the device is turned on for 10min The nasal cannula...

Page 8: ...est choice Please select the suitable humidifier bottle for your device and fix it on the machine as required Do not operate under a revolving situation Refer to the humidifier bottle connection indic...

Page 9: ...p Put Double insulated Avoid the rain Keep dry Type BF applied part Fragile handle with care 220V AC 220V Forbidden to sit Non ionizing electromagnetic radiation Waste Electrical and Electronic Equipm...

Page 10: ...er If the protective cover is damaged please check whether the product is damaged or not immediately Then check if there are any missing parts or accessories according to the packing list below Packin...

Page 11: ...f the air flow and ensur e the normal heat 22 Connect the humidifier bottle and operate it as following instruction3 a Fill the humidifier bottle with distilled water as instructed DO N OT fill more o...

Page 12: ...ctly 52 Adjust the output flow of oxygen according to the needs of patients Turn the flow adjustment knob counterclockwise to increase the flow and turn it clockwise to decrease the flow5 Insert the n...

Page 13: ...ieve and a pressure swing adsorption process When the device is operating under room temperature the oxygen in the air can be separated and transform into highly concentrated oxygen to meet the medica...

Page 14: ...ature environment for over four hours before use ADDITIONAL REMARKS Classification of protection against electric shock3Class II Classification according to the degree of protection against electric s...

Page 15: ...ltage3 220V 260Hz4 Working3Continuous working4 With the application of oxygen free discharge part of the protective effect of defibrillation4 Oxygen Concentrator is for non permanent installation equi...

Page 16: ...Range of adjusting flow 0 5L min adjusted continuously Net Weight Kg 14 3 Noise dB A 47 Dimension mm L380 W249 H593 Oxygen output pressure 50kPa 5kPa Yellow low oxygen indicator light Refer to 15 c W...

Page 17: ...ttle 4 Product label product performance product serial number 5 Air filter cover contains air filtering device and air filter foam 6 Air filter foam3To prevent dirt dust fibers into the machine 7 Hea...

Page 18: ...ster flow to control the flow to go faster or slower A B C B D E FC C B G D H II B D GI J K L BM F D NAO B GI J PP LMCBO D M P LQF CBF LRP C BF5 STD UV 1 F LC F FREE L D GI J RE WXKCB5YU UT F D GI J Z...

Page 19: ...600 minutes l Cumulative timing3Digital show for 6 digits n Total operating hours after the machine leaves the factory n When the machine is turned on or the timer function is cancelled the total runn...

Page 20: ...d low pressure occur at same time the machine will light up the Red Alarm Indicator issue an alarm and display Lo Po2 on the LED display The Oxygen Concentrator will stop working within one minute Ple...

Page 21: ...gen Concentrator are considered low priority The alarm system of the Oxygen Concentrator is set before leaving the factory and users CANNOT change the settings of the alarm system CLEANING AND MAINTEN...

Page 22: ...s repairment or replacement are provided at its option the defective material or part This warranty does not cover damage caused by accident misuse abuse alteration and other defects not related to ma...

Page 23: ...ator circuit breaker activated Press the concentrator circuit breaker button if equipped located below the power switch Use a different wall outlet if the situation occurs again B Unit operates within...

Page 24: ...ake sure there is nothing restricting the unit exhaust F Red Service Required light is on and an intermittent audible signal is sounding 1 Flow meter is not properly yet Ensure the flow meter is prope...

Page 25: ...F emissions CISPR 11 Group 1 The ME uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF...

Page 26: ...output lines 2kV for power supply lines Not applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV line s to line s 2kV line s to earth...

Page 27: ...e customer or the user of the ME should assure that is used in such and environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6...

Page 28: ...radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be consid...