![FACIOX OC503-BL User Manual Download Page 21](http://html1.mh-extra.com/html/faciox/oc503-bl/oc503-bl_user-manual_536679021.webp)
OXYGEN CONCENTRATOR USER MANUAL
"0!
If the air temperature of the oxygen outlet is over 46 ° C, the machine
will light up the Red Alarm Indicator, issue an alarm and the display "E07" on
the LED display. The machine will stop working within one minute. Please turn
off the power and contact your local service or manufacturer.
If the oxygen outlet or pipeline is blocked, the machine will light up the
Red Alarm Indicator, issue an alarm and the display "E08" on the LED display.
Please turn off the power; check the humidifier bottle and tube if it is blocked.
If either one of them is blocked, please check if it is possible to clean it. If not,
please contact your local service or manufacturer.
Note: All alarms of this Oxygen Concentrator are considered low priority.
The alarm system of the Oxygen Concentrator is set before leaving the factory,
and users CANNOT change the settings of the alarm system.
CLEANING AND MAINTENANCE INSTRUCTIONS
1
2
Clean the outside shell
3
one
/
twice per month. Turn off the power and
disconnect from all power source before cleaning. Clean the device
exterior using a wet towel with a mild household cleaner and wipe it dry.
2
2
Clean the humidifier bottle
3
Use detergent and hot water to clean the
humidifier bottle separately, or use white vinegar with water at the ratio
of 1:3 mixtures of fungicides. Soak the humidifier bottle in the solution
for 30 minutes and wipe it dry.
3
2
Clean the air filter foam
3
take off the filter foam on the back side of the
device exterior
*
clean it with detergent and rinse thoroughly with clean
water. Dry in the air. Please don’t use it if it is not dry thoroughly. This is