![FACIOX OC503-BL User Manual Download Page 20](http://html1.mh-extra.com/html/faciox/oc503-bl/oc503-bl_user-manual_536679020.webp)
OXYGEN CONCENTRATOR USER MANUAL
"1!
When the oxygen concentrator is working, it will automatically monitor
the machine status.
If the oxygen purity is below 82% then the machine will light up the
Yellow Low Purity Indicator, issue an alarm and display the “Lo-o2” on LED
display. Please turn off the power and contact your local service or the
manufacture.
If a low pressure occurs, the machine will light up the Red Alarm
Indicator, issue an alarm and display "Lo-P" on the LED display. The Oxygen
Concentrator will stop working within one minute. Please turn off the power
and contact your local service or the manufacturer.
If both cycle error and low pressure occur at same time, the machine will
light up the Red Alarm Indicator, issue an alarm and display "Lo-Po2" on the
LED display. The Oxygen Concentrator will stop working within one minute.
Please turn off the power and contact your local service or manufacturer.
If a high pressure occurs, the machine will light up the Red Alarm
Indicator, issue an alarm and display "Hi-P" on the LED display. The Oxygen
Concentrator will stop working within one minute. Please turn off the power
and contact your local service or the manufacturer.
If both cycle error and high pressure occur at same time then the
machine will light up the Red Alarm Indicator, issue an alarm and display
"Hi-Po2" on the LED display. The Oxygen Concentrator will stop working
within one minute. Please turn off the power and contact your local service or
manufacturer.