Crossover™ Manual V1/62920016 ©Fabtech Systems LLC. All Rights Reserved.
Alignment and Initial Ambulation Settings
This procedure addresses both ambulation and sport
When setting up the Crossover™ knee for the first time on a new patient, the goal is to align the knee in a sta-
ble condition for ambulation. All adjustments to tune the knee are after the patient is set in a stable alignment.
Make sure the patient is in parallel bars for safety as the knee will be in free swing with no or very little re-
sistance in the initial set up phase.
1.
Unpack the box and make sure all items are in the package.
2.
Remove air cap on shock and remove any residual air pressure. Depress valve to remove air.
(
Figure 1)
3.
Set all shock settings to the zero position. (Page 11-12)
4.
Remove protective covers for easy adjustments.
5.
Attach Crossover™ to prosthetic.
6.
For recommended beginning alignment.
(Figure 2)
Warning! If the patient is not stable they can fall at this stage of the fitting.
Figure 1
Figure 2
9