5
Initialization of the Columns and
Tilt Actuator
For functions on the columns and the tilt actuator
to work correctly, an initialization should be carried
out the very first time the EZ1000 is used.
To begin the initialization, first check that the
emergency stop is not activated. Then follow the
below steps:
1.
First press the DOWN button on the hand
control to lower the 2 columns to their lowest
position. When both columns are in their
lowest position, a faint “click” sound will be
heard from the control box.
2.
Release the DOWN button for a few seconds.
3.
Then once again press the DOWN button and
hold the button for three seconds.
4.
Press the TILT BACK button on the hand
control to retract the actuator and tilt the
chair backwards to its rear-most position.
5.
Visually check that the chair is tilted all
the way back and that the tilt actuator is
completely retracted. When the tilt actuator is
completely retracted and the chair has reached
its rear-most position, a faint “click” sound will
be heard from the control box.
6.
Release the TILT BACK button for a few
seconds.
7.
Then press the TILT BACK button again and
hold the button for three seconds.
8.
The initialization is now complete and you can
start using the EZ1000.
Height Control
Secure the resident with the safety belt at all times.
Raise and lower the EZ1000 with the hand control.
•
Raise the EZ1000 to desired height by pressing
the up button on the hand control.
•
Lower the EZ1000 by pressing the down button
on the hand control.
Emergency Stop Button
Recline/Tilting the Chair
•
Press the tilt arrow button on the hand control
until required reclined position (max 25
degrees) is achieved.
•
When leveling, press the opposite arrow
button. The tilt will stop in neutral position.
•
The chair can be tilted a few degrees forward to
ease transfer.
NOTE: The forward tilt should always and
only be performed when the chair is in its
lowest position.
Load limit: 440 lbs.
Raise the EZ1000 to a convenient transport height
for both resident and caregiver when transporting
to the shower area and while showering. Recline/tilt
the chair to a convenient position for both resident
and caregiver when required (the caster brakes
should be locked).
Emergency Stop Button
The EZ1000 is equipped with a red emergency
stop button that can be activated if an unexpected
incident should occur. Push to activate and twist
clockwise to reset.
Hand Control
Column/Height
Adjustment
Buttons
Recline/Tilt
Adjustment