- 33 -
4. Humidification
1. Start of manual defrost operation (Forced defrosting)
Stop the control operation or turn on [Manual Drain]
(forced draining) key switch during control operation.
If you press [Defrost] key, the setting temperature will be
at 25
℃
regardless of operation mode, and forced defrost
operation will start working automatically.
- Refrigerator lamp illuminates.
- Defrost Running lamp illuminates.
- Temp. Heater lamp illuminates.
* Depending on the temperature inside the chamber.
Empty
Full
Tank Water Level
Level 1 (Approx. 2L)
Level 2 (Approx. 6L)
Level 3 (Approx.10L)
2. Performance of defrost operation
When setting temperature at 25
℃
for defrost
operation and the operation is continued for 20
minutes, it stops automatically and the temperature
inside the chamber returns to the setting temperature
that was active when the control operation was
stopped, or control operation before defrosting will
restart.
If the defrost operation is executed immediately after
the refrigeration unit stops, you need to wait 80
seconds until it restarts and hot gas flows into it. This
is a timer for protection of refrigeration unit.
*Even if the temperature inside the chamber does
not reach to 25
℃
(defrost temperature),
the operation stops automatically after elapsing
20 minutes.
3. Manual stop of Defrost operation
If you would like to stop manually while
operating defrost, press [Defrost] key again.
Turn on [Water Supply] key.
The water is supplied into the internal tank from
the external tank until it is full. (Approx. 10L)
Check the water level of the internal tank.
See the indicator of [Tank Water Level] on the
control panel to check the supplying status.
The indicator [Tank Water Level] is changed
from [Empty] (red) to level 1
~
3 and to [Full]
(green) illumination in proportion as increasing
of water.
Turn on [Run Stop] key.
When the water level reaches to Level 1 even if
it is not full, the solenoid pump works and the
water is supplied into the top and bottom
humidification tray. and the wick holder.
About 10 minutes after turning on [Run Stop]
key, the water is supplied into the top and
bottom humidification tray and the wick holder.
While draining (Manual Drain key is turned on.)
or when the setting temperature is lower than
5
℃
, the water is not supplied into the top and
bottom humidification tray and the wick holder.
Except above conditions and when the setting
humidity is 0.0%RH, the humidification control
is not executed although the water is supplied
into the top and bottom humidification tray and
the wick holder.
If once the level reaches to 3 (or Full) and it
lowers to the level 1, the water is supplied
automatically from the external tank to the
internal tank.
If the water level raised to the level 3 (or Full)
from 2, turn off [Water Supply] key once, and
turn on it again.
(1)
(2)
(3)
*
*
*
In case the solenoid pump does not work.
The hose of supplying side may include air.
Open the ventilation cover and squash the hose
during the operation to remove air.
Internal tank
Squash to remove air.
To the suction
port of pump
Socket
The required time for filling water into the
internal tank depends on the position of
external tank (whether it is set on the right or
left side) or layout of hose connection.
Required time until the internal tank is full (10L).
- In case the external tank is
set on the observer’s right
- In case the external tank is
set on the observer’s left.
Approx. 20 minutes
Approx. 35 minutes
When supplying water into the top and bottom
humidification tray and wick holder, you will hear
the operating noise of electro magnetic pump and
solenoid valve. However, it is not a malfunction.
Also, you will hear the operating noise of solenoid
valve when selecting cooling capacity of
refrigerator, and while defrosting operation is active.