3-4 Overview of operation
Program mode
1. Program 1 (Auto start program)
You can set a temperature, humidity and time of
onset. After the setting time elapses, operation
starts and continues at the set temperature.
* Time setting range : 1 minute
~
99 days 23
hours 59 minutes.
Normal mode
・
Set-point control operation
* Temperature setting range
-15.0
~
85.0
℃
(Input range -20.0
~
90.0
℃
)
Humidity setting range 25.0
~
98.0
%
RH
(Input range 0
~
98.0
%
RH)
・
Program and normal mode share the same temperature setting range.
・
Refer to the humidification control range on P.4 to
set humidity value.
・
Although the humidity input range is 0
~
98.0
%
RH,
the actual control operation starts form 2%RH.
Time
Temperature
Humidity
Temperature
Humidity
Time
Temperature
Humidity
Temperature
Humidity
Set time
Time
Temperature
Humidity
Temperature
Humidity
Set time
Two operation modes, which are normal and program
mode, are available. In program mode, there are five
patterns of program from No.1 to 11.
For more detail about the setting procedure, please
refer to the key operation manual.
After setting a temperature and starting the operation,
the unit controls the temperature at the set point.
2. Program 2 (Auto stop program)
You can set a temperature, humidity and stop
time. After the setting time elapses, operation
stops.
* Time setting range : 1 minute
~
99 days 23
hours 59 minutes.