4-5
Utility connection
12
Inappropriate connection may cause a fire or
electric shock hazard.
Check the voltage, phase and
capacity of power source before
connecting.
WARNING
Over-current may cause cable burn, fire.
Do not use the branching
socket or table tap.
WARNING
Model
Required power source
Voltage
Capacity
CA-1330
AC 220V
Single phase
15A
CAUTION
Remove the dirt on the grounding
adapter,
outlet and mains connector.
Dirt on these parts may cause tracking or fire.
WARNING
The unit must be earthed.
Without grounding the unit, the product may
cause electric shock hazard.
(
1
)
Check the voltage, phase and capacity of
the power source. Required power source
is as shown in right table.
(
2
)
Check the type of outlet at the installation site.
(
Do not connect the mains connector yet.
)
If the outlet has earth terminal, mains
connector can be connected.
Length
Thickness
(O.D.)
Mains plug
Cable
cross-
sectional
area
2.8m
About
9mm
3-pronged
with earth
terminal
1.5mm
2
Do not use branching socket when connecting to
power source.