The exhaust temperature of the
unit may exceed 60
℃
in certain
cases.
To keep the best performance of the product, leave
the space between product and wall, ceiling plane.
The distance between the product and wall, ceiling
plane must be longer than the one mentioned in the
right picture.
※
Keep a distance of 3 cm or more between
the unit backside and the wall surface and 3
cm or more between the unit top and the
ceiling.
※
The unit can be used with one of left or right
side blocked. In such an event, be sure to
keep a distance of 40 cm or more between
one of unit sides and the wall surface.
(※
For use at the room temperature of
30
℃
or below, the unit can be used if a
distance of 10 cm is secured between one
side and the wall surface.
)
※
Install the unit so that its front (with a
ventilation port) does not come under the
table.
(
See the table on P.10
)
※
With the unit installed with its side covered
40 cm under the table, the cooling capacity
is degraded by maximum 14% as compared
with the case of installation outside the
range of table.
At the -20
℃
setting, the cooling capacity is
degraded by maximum 25%.
CAUTION
CAUTION
4-2
Installation condition
4-1
Installation environment
Installation
●
No direct sun light
●
Ambient temperature must be from 5 to 35
℃
.
●
Well-ventilated.
●
No inflammable solid or liquid or gas around the unit.
●
No dew condensation
●
Lesser humidity and no dripping on the unit.
●
Lesser dust
●
Level and stable
(
Check the weight of the product during operation.
)
●
Indoor use only
Do not place anything on top of the
product.
CAUTION
Keep enough space around the unit.
Be careful about the environment of
installation. In particular, pay due
attention on the location, air
conditioning, and ventilation.
CAUTION
Since air-cooled type refrigeration unit is
equipped with this product, heat is exhausted from
the unit.
Select the installation site that can be ventilated
well so that the ambient temperature won’t rise
because of exhausted heat. Using the product in
high ambient temperature may worsen the
operation efficiency or cooling capability. Also,
refrigeration unit will be hot and operate under
high pressure, which may cause malfunction.