User Manual MUT_RD_009H_GB Page
36
of 43
3.
Once the first session of treatment done, release the firing pedal
For 6-sectors protocol
, refers to
Step 6
“Treatment completion”.
For 8-sectors protocol:
after a loading period, a
message is displayed, asking the user to rotate the
therapy probe in the coupling.
4.
Rotation of the therapy probe in the
coupling cone
(only for 8-sectors protocol):
Rotate the therapy probe in the coupling cone in
the clockwise direction up to the stop, without
forcing, by using the wings of the probe. Once
done, release the wings in order to maintain the
probe in its second position. Then, push the
OK
button.
Do not move the coupling cone, and keep it
centered on the eye during the rotation of
the therapy probe.
5.
Start the second treatment session by
releasing
(if not already done) then by
pressing - and maintaining pressure - on the
firing pedal.
The shoot on the first sector of the sequence
starts after a short countdown.
The firing time on each sector lasts
8 seconds
,
and there is a 20-second interval between two
consecutive shots.