3.4.2.3
Sheet of detachable labels
The individual box contains a sheet of 4 identical labels to be used by the operator/user in order to
ensure traceability of the product according to the procedures in force in the establishment. The
information contained is similar to that present on the adhesive label affixed on the box lid.
3.4.3
Electromagnetic compatibility tables (EMC)
The medical device (EyeOP1 control unit and associated EYEOP-PACK consumables) requires specific
precautions to be taken with respect to the EMC and must be installed and operated according to the EMC
information provided in the tables below.
Cellular phones and other mobile communication devices must not be used close to the medical device,
since they could affect its operation. Any other device used close to the medical device must comply with
the IEC 60601-1-2 standard. Any incident related with electromagnetic compatibility must be reported to
the manufacturer.
Exposure of patients equipped with active implantable medical devices (pacemaker…) to the EyeOP1
control unit and associated EYEOP-PACK consumables must be supervised by a cardiologist.