User Manual MUT_RD_009H_GB Page
28
of 43
Before starting the procedure, make sure you have read and understood the ‘safety and
precautions for use’ part of
Section 3
.
The anesthesia (general, local or topical) must be performed on the patient only after the medical
device has been switched ON and after checking its proper operation.
Leave enough time for the anesthesia to be effective. The treatment should not be painful if the
anesthesia is administered properly.
The patient must be immobile and lying flat on his/her back, so as to have a vertical optical axis
(
see below
) for the entire duration of the treatment and preparation.
Before each treatment, make sure that the printer is fed with enough paper in order to print the
treatment report: see ‘
§4.3. Checking,
replacing and repositioning the printer paper’
of
Section 7.
5.3.1
Step 1
.
Information for the physician-ophthalmologist-user
After turning on the machine, the user must identify
himself in the Main Menu:
1.
Select
USER
If the previous treatment procedure was
performed by the same user, you can select the
key with the latter’s last and first
names, and
thus go directly to step 2.
2.
Enter the user’s last name and validate.
3.
Enter the user’s first name and validate.
The physician’s first name and last name c
an
include between 1 and 20 characters each, and
must not begin with a space “
˽
”.
4.
Once the two fields have been correctly filled in, the
system automatically moves to the next step.
5.3.2
Step 2
.
Patient information