ExTox ET-4D2-24 Instruction Manual Download Page 12

 

Instruction Manual

ExTox IMC-8/-4

 
 

Page 12 of 22 

 

BA_IMC-8_-4_e_2019-12-17.docx 

 

IMC-8D (IMC-4D) 

Application 

Digital Input E3 

Status signal housing fans (

only allocated in systems with Flow-Meter, 

see 6.1.2

Digital input E4 

External activation of maintenance mode is not provided 

 

5.2.4

 

Calibration and Adjustment 

Calibration and adjustment of the biogas-types is only done via automatic control. 

 

Preparation 

You have to enter the concentrations of the used calibration gases in the menu "System Param-
eter/Mode/Bio/Timer". On the right side of the display you could enter the concentrations for 

CH

4

, O

2

, CO

2

 and H

2

S. These values are necessary for automatic calibration. You will normally 

find this information on the test gas bottle. In case the test gases are filled in different pressure 

gas bottles, you have to follow the instructions following below for all used test gases. For cali-
bration of the O

2

-Measuring Channel you have to apply a test gas which does not contain any 

oxygen. 

 

Course 

You have to select the menu point "Calibration" 'in the menu "Setup". For that purpose you have 

to activate minimum Access Level 2 before. The calibration menu appears and the internal mag-
netic valve switches to flushing air. In case the system should not be calibrated you could leave 

the menu via "Esc". By means of "Start" the automatic calibration is started. The analogue out-
puts keep the measured values of the last measurement. 
Calibration is done in five phases:  

Phase  Indication on Display  

ET-8D-(ET-4D2-) 

Course 

1 RINSING 

AIR 

(RINS AIR) 

The system is flushed with air for three minutes 

READING ZERO 
(READ ZERO) 

The gas suction pump is switched off. The system waits one 
minute until stabilisation of measured values. Then the 

measured values are stored (zero point values for CH

4

, CO

2

H

2

S or 20.9 Vol.-% for O

2

). 

3 RINSING 

GAS 

(RINS GAS) 

The gas suction pump is switched on again and the system 
is flushed with connected test gas for four minutes.  

Attention: 

Only during this phase the test gas bottle has 

to be connected and the test gas has to be applied to the 
IMC. The test gas has to be applied pressure less. It has 
always to be ensured that an adequate test gas volume 

stream is provided, that means on application of test gas 
has to be set which corresponds to the one in measuring 

operation

4

. This is to avoid that air is additionally sucked in 

and dilution of the test gas resulting from this falsifies cali-
bration / adjustment. 

READING SENSITIVITY 

 

(READ SENS) 

The measured gas pump is switched off and batch meas-
urement is done within two minutes. 

                                          
 

4

 

Depending on type the necessary test gas volume stream may be considerably above the indicated volume stream.

 

Summary of Contents for ET-4D2-24

Page 1: ...Instruction Manual ExTox Integral Measuring Concepts IMC 8 and IMC 4...

Page 2: ...trial application The Integral Measuring Concepts serve for monitoring gas concentrations in sealed off processes or not accessible areas The IMC Series combines all necessary components in one compac...

Page 3: ...5 3 2 Calibration and Adjustment 16 5 4 Type M 17 5 4 1 Description 17 5 4 2 Replacement of a Transmitter Module 17 6 Operation of the IMC 18 6 1 Flow rate monitoring 18 6 1 1 Types with DDL Module 3...

Page 4: ...itten documentation and is on delivery attached Type specific data for each device should be taken from a corresponding separate and specific Data Sheet This Data Sheet forms consequently part of this...

Page 5: ...ntegrate controlling and monitoring of the components for sampling and preparation of measured gas The measuring channels which are not used for analysis can easily be taken for external trans mitters...

Page 6: ...easuring points temporally one after another Supplementation of switching between measuring points forms a cost saving alternative to using several IMC One single IMC performs the analysis one after a...

Page 7: ...rms and messages as well as keys is installed in the door of the wall mounted housing For description of the indications and facilities please see the Instruction Manual of the control unit In the sur...

Page 8: ...n the menu Channel Configuration by setting the Mode ON Status Monitoring The following status monitoring functions are integrated in this type Rotary speed monitoring of both housing fans Flow rate m...

Page 9: ...for switching between measured gas and test gas Digital input E2 Status signal flow rate only allocated in systems with Flow Meter see 6 1 2 Digital input E3 Status signal housing fans only allocated...

Page 10: ...during measuring phase Attention This means that reset will be pos sible in the next phase If necessary measurement can be started manually see above to reduce the waiting time on longer cycle times T...

Page 11: ...ith DDL Module see 6 1 1 Lower limit of flow rate monitoring in l h Under scale leads to system fault Standard setting 15 l h FLOW MAX 0 to 300 l h Only activated on systems with DDL Module see 6 1 1...

Page 12: ...ystem should not be calibrated you could leave the menu via Esc By means of Start the automatic calibration is started The analogue out puts keep the measured values of the last measurement Calibratio...

Page 13: ...valves which access the corresponding measuring line for measurement inside the IMC are driven by the control unit Via relay outputs the control unit signalises the measuring point for which the measu...

Page 14: ...alve for measuring point 4 activated Relay 9 closed Magnetic valve for measuring point 5 activated IMC 8 Relay 11 closed Magnetic valve for measuring point 6 activated IMC 8 Relays K12 K18 K19 K6 K7 K...

Page 15: ...witching between 4 measuring points OPTIONS M1 ON 1 H RS232 1 M2 ON 1 H FLOW MIN 15 l h M3 ON 1 H FLOW MAX 300 l h M4 ON 2 L MAX CH 4 v ESC SAVE MAX CHANNEL indicates the number of measuring points he...

Page 16: ...cognized by the software and calibration menu is automatically extended CALIBRATION FLOW 31 l h F1 CAL H2S 0 3000 ppm F2 CAL H2S 0 100 ppm H2S HI H2S LO ESC In case calibration is started with F1 H2S...

Page 17: ...l get a Replacement Transmitter Module Art 295xxx on which all wear parts have been replaced The module has been checked completely and adjusted It has the same warranty as for a new module In case yo...

Page 18: ...every second In case the serial data link between control unit and DDL Module is interrupted for more than 15 s the message COMERROR is issued A fault of control unit is indicated at the control unit...

Page 19: ...midification is done by means of a Peltier cooler The temperature of measured gas is factory sided adjusted to 5 C and monitored During the warm up phase a fault message is issued until the cooler rea...

Page 20: ...bigger climatic fluctuations In general we do not recommend an installation outside as in principle measuring accuracy and lifetime are influenced in a negative way In case it is however necessary pl...

Page 21: ...efore the line gets into the IMC If necessary please contact your partner at ExTox Operation in most of the compositions of measured gas is reliably possible due to the construction of the IMC Only fo...

Page 22: ...O2 H2S3000 295008 Replacement Transmitter Module CH4 O2 H2S100 940006 Measured Gas Pump 940007 Condensate Trap 940010 Hose Pump 300 ml h 940011 Replacement Hose for hose pump 300 ml h 940021 Measured...

Reviews: