ExTox ET-4D2-24 Instruction Manual Download Page 10

 

Instruction Manual

ExTox IMC-8/-4

 
 

Page 10 of 22 

 

BA_IMC-8_-4_e_2019-12-17.docx 

 

5.2

 

Type Biogas 

5.2.1

 

Functions 

 

Course of Biogas-Measurement 

During operation mode "BIO" discontinuous measurement takes place. All channels for biogas 
measurement are selected in the menu Channel Configuration by setting the mode TIMER. Three 
phases are run through cyclically: 
1.

 

Air: the gas flow is flushed with air that means the magnetic valve has switched to the inlet 
test gas / flushing air. 

2.

 

Flushing: the system switches to the measured gas inlet. The gas concentration is not meas-

ured during this time. This phase is needed to transport the sampled gas from the process to 
the IMC using the gas flow. The gas suction pump is switched off two minutes before end and 
the gas flow is closed. Measurement in batch mode is prepared. 

3.

 

Measuring: the system measures the actual values when changing over to phase measure-
ment. During this phase the gas suction pump remains switched off.  

 
You can manually interfere in this process. Measurement can be started by pressing simultane-

ously the keys F8+F3 (IMC-8D) or F6+F3 (IMC-4D). The IMC-Biogas first changes to the phase 
flushing and starts the measurement later on.  
For measuring points in TIMER Mode all measured values, alarms and outputs remain frozen on 
the latest value of the previous measuring phase during the phases “air” and “flushing”. Latching 
alarms can only be reset during measuring phase. 

Attention:

 This means that reset will be pos-

sible in the next phase. If necessary measurement can be started manually (see above) to reduce 
the waiting time on longer cycle times.   
The times for the individual phases can be adjusted in the menu Timer which can be reached via 
the menu System Parameter and selection of the menu setting Mode BIO. 

 

Monitoring of ambient air 

Channels which are not needed for the IMC measurement can easily be used for a continuous 

monitoring of ambient air. Performance of these channels corresponds to the normal measuring 
operation of the control unit. Channels for monitoring of ambient air are also activated in the 
menu Channel Configuration by setting the Mode ON. 

 

Status Monitoring 

The following status monitoring functions are integrated in this type: 

 

Rotary speed monitoring of both housing fans 

 

Flow rate monitoring: message for over- and under-scale of nominal range 

 

Monitoring Tmax or 

T of the measured gas cooler (on request) 

In the first two cases a message leads to system fault of the control unit. 
As long as the measured gas cooler does not reach its operating temperature, for example during 
the warm-up phase after initial operation, the gas suction pump remains switched off. 

5.2.2

 

Parameter Settings 

 

System Parameter  

In the menu System Parameter the operation mode „BIO“ instead of „ET-8“ and „ET-4“ is factory-
sided activated.   

Attention

Changing of these settings should in no way be done by the user. Changed measuring 

functions may otherwise lead to loss of safety functions. 
In the submenu BIO the following changes can be performed. 

Summary of Contents for ET-4D2-24

Page 1: ...Instruction Manual ExTox Integral Measuring Concepts IMC 8 and IMC 4...

Page 2: ...trial application The Integral Measuring Concepts serve for monitoring gas concentrations in sealed off processes or not accessible areas The IMC Series combines all necessary components in one compac...

Page 3: ...5 3 2 Calibration and Adjustment 16 5 4 Type M 17 5 4 1 Description 17 5 4 2 Replacement of a Transmitter Module 17 6 Operation of the IMC 18 6 1 Flow rate monitoring 18 6 1 1 Types with DDL Module 3...

Page 4: ...itten documentation and is on delivery attached Type specific data for each device should be taken from a corresponding separate and specific Data Sheet This Data Sheet forms consequently part of this...

Page 5: ...ntegrate controlling and monitoring of the components for sampling and preparation of measured gas The measuring channels which are not used for analysis can easily be taken for external trans mitters...

Page 6: ...easuring points temporally one after another Supplementation of switching between measuring points forms a cost saving alternative to using several IMC One single IMC performs the analysis one after a...

Page 7: ...rms and messages as well as keys is installed in the door of the wall mounted housing For description of the indications and facilities please see the Instruction Manual of the control unit In the sur...

Page 8: ...n the menu Channel Configuration by setting the Mode ON Status Monitoring The following status monitoring functions are integrated in this type Rotary speed monitoring of both housing fans Flow rate m...

Page 9: ...for switching between measured gas and test gas Digital input E2 Status signal flow rate only allocated in systems with Flow Meter see 6 1 2 Digital input E3 Status signal housing fans only allocated...

Page 10: ...during measuring phase Attention This means that reset will be pos sible in the next phase If necessary measurement can be started manually see above to reduce the waiting time on longer cycle times T...

Page 11: ...ith DDL Module see 6 1 1 Lower limit of flow rate monitoring in l h Under scale leads to system fault Standard setting 15 l h FLOW MAX 0 to 300 l h Only activated on systems with DDL Module see 6 1 1...

Page 12: ...ystem should not be calibrated you could leave the menu via Esc By means of Start the automatic calibration is started The analogue out puts keep the measured values of the last measurement Calibratio...

Page 13: ...valves which access the corresponding measuring line for measurement inside the IMC are driven by the control unit Via relay outputs the control unit signalises the measuring point for which the measu...

Page 14: ...alve for measuring point 4 activated Relay 9 closed Magnetic valve for measuring point 5 activated IMC 8 Relay 11 closed Magnetic valve for measuring point 6 activated IMC 8 Relays K12 K18 K19 K6 K7 K...

Page 15: ...witching between 4 measuring points OPTIONS M1 ON 1 H RS232 1 M2 ON 1 H FLOW MIN 15 l h M3 ON 1 H FLOW MAX 300 l h M4 ON 2 L MAX CH 4 v ESC SAVE MAX CHANNEL indicates the number of measuring points he...

Page 16: ...cognized by the software and calibration menu is automatically extended CALIBRATION FLOW 31 l h F1 CAL H2S 0 3000 ppm F2 CAL H2S 0 100 ppm H2S HI H2S LO ESC In case calibration is started with F1 H2S...

Page 17: ...l get a Replacement Transmitter Module Art 295xxx on which all wear parts have been replaced The module has been checked completely and adjusted It has the same warranty as for a new module In case yo...

Page 18: ...every second In case the serial data link between control unit and DDL Module is interrupted for more than 15 s the message COMERROR is issued A fault of control unit is indicated at the control unit...

Page 19: ...midification is done by means of a Peltier cooler The temperature of measured gas is factory sided adjusted to 5 C and monitored During the warm up phase a fault message is issued until the cooler rea...

Page 20: ...bigger climatic fluctuations In general we do not recommend an installation outside as in principle measuring accuracy and lifetime are influenced in a negative way In case it is however necessary pl...

Page 21: ...efore the line gets into the IMC If necessary please contact your partner at ExTox Operation in most of the compositions of measured gas is reliably possible due to the construction of the IMC Only fo...

Page 22: ...O2 H2S3000 295008 Replacement Transmitter Module CH4 O2 H2S100 940006 Measured Gas Pump 940007 Condensate Trap 940010 Hose Pump 300 ml h 940011 Replacement Hose for hose pump 300 ml h 940021 Measured...

Reviews: