Extech Instruments Oyster-10 User Manual Download Page 8

 

Oyster10-en-US_v2.3   9/15 

8

Maintenance and Troubleshooting 

Electrode Cleaning 

The type of electrode coating that occurs over time will determine the cleaning method. Soft 
coatings can be removed by vigorous stirring or by using a squirt bottle. Organic chemical coating 
(hard coatings) should be chemically removed. Only in extreme cases should the bulb be cleaned 
mechanically. 

Battery Replacement 

Install a new battery when the low battery indicator appears on the LCD. 

 

1. 

Open the calibrator’s lid as far as possible.  

2. 

Open the battery compartment using a coin at the arrow indicator. 

3. 

Replace the battery and close the cover. 

 

 Battery 

Safety 

Reminders 

  Please dispose of batteries responsibly; always observe local, state, and federal regulations 

with regard to battery disposal. 

  Never dispose of batteries in a fire. Batteries may explode or leak. 

  Never mix battery types. Always install new batteries of the same type.

 

Troubleshooting the Meter

 

Meters are the least likely cause of pH measurement problems. To rule out the meter check the 
following: 
1.  Check the 9V battery and replace it if necessary. 
2.  When using the AC adaptor for power, check that the line cord is firmly connected to the meter. 
3.  Replace the existing electrode connector with optional BNC shorting plug (shorts the ground 

ring to the center pin of the meter’s BNC jack) and check the following: 
a)  Set the measurement mode to pH. The LCD reading should be stable. 
b)  Adjust the calibration control for an LCD reading of 7.00. 
c)  Change to the mV measurement mode. The LCD should read 0 mV. 
d)  Note that an out of range reading is indicated by dashes ( -------). 

www.ponpe.com

contact:
[email protected]

Summary of Contents for Oyster-10

Page 1: ...USER GUIDE pH mV Temperature Meter Model OYSTER 10 www ponpe com contact sale ponpe com...

Page 2: ...lude diagnostic displays MIN MAX readings audible alerts 5 point buffer recognition Auto Manual temperature compensation This instrument is shipped fully tested and calibrated and with proper use will...

Page 3: ...calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 Open the top half of the meter to an appropriate viewing position 3 Power the meter with the...

Page 4: ...s complete unplug the electrode and rinse it in distilled water 8 Place approx 1 ml of pH 4 solution in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required...

Page 5: ...e MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This procedure is for...

Page 6: ...er or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible 7 When calibrat...

Page 7: ...ure readings LCD viewing angle 90 o to 180 o Over range Indication Displays Power 9 volt battery or optional AC adapter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery ico...

Page 8: ...nd federal regulations with regard to battery disposal Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of the same type Troubles...

Page 9: ...be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Long term storage wi...

Page 10: ...mages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibrat...

Page 11: ...it totale de FLIR est limit e la r paration ou au remplacement du produit La garantie d finie ci dessus est inclusive et aucune autre garantie crite ou orale n est exprim e ou implicite Calibrage r pa...

Page 12: ...entes La responsabilidad total de FLIR est limitada a la reparaci n o reemplazo del producto La garant a precedente es inclusiva y no hay otra garant a ya sea escrita u oral expresa o impl cita Servic...

Reviews: