MAX/MIN Display 

1.  The meter memorizes the highest (MAX) and lowest (MIN) readings when prompted. 
2.  Press and hold the MAX/MIN button until both the 

MAX

 and 

MIN

 icons appear on the LCD. The 

meter will now begin keeping track of the MIN and MAX readings. 

3.  To view MAX/MIN values, press the MAX/MIN button again; the meter automatically displays the 

highest and lowest readings in succession and then returns to the MAX/MIN recording mode. 

4.  To exit the MAX/MIN mode, press and hold the MAX/MIN button until the 

MAX

 and 

MIN

 icons 

disappear. 

5.  To clear the MAX/MIN memory, cycle the meter power. 

External Temperature Offset Adjustment 

The Offset Adjustment is used to fine tune the temperature reading of the external temperature 
probe. This procedure is for the external probe only when ATC is active. 
1.  Place the temperature probe into a known temperature and note any temperature reading error. 
2.  Simultaneously Press and Release the 

and 

buttons, 0.0 will appear on the LCD 

temperature reading  

3. .Press 

the 

 or 

 button to set the display equal to the error value determined in step 1. 

4.  Press the STORE button to save the offset value into memory, 
5.  Test the temperature probe in the know temperature source to verify the correction was proper 

and the adjustment made in the correct direction. 

www.ponpe.com

contact:

[email protected]

Summary of Contents for Oyster-10

Page 1: ...USER GUIDE pH mV Temperature Meter Model OYSTER 10 www ponpe com contact sale ponpe com...

Page 2: ...lude diagnostic displays MIN MAX readings audible alerts 5 point buffer recognition Auto Manual temperature compensation This instrument is shipped fully tested and calibrated and with proper use will...

Page 3: ...calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 Open the top half of the meter to an appropriate viewing position 3 Power the meter with the...

Page 4: ...s complete unplug the electrode and rinse it in distilled water 8 Place approx 1 ml of pH 4 solution in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required...

Page 5: ...e MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This procedure is for...

Page 6: ...er or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible 7 When calibrat...

Page 7: ...ure readings LCD viewing angle 90 o to 180 o Over range Indication Displays Power 9 volt battery or optional AC adapter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery ico...

Page 8: ...nd federal regulations with regard to battery disposal Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of the same type Troubles...

Page 9: ...be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Long term storage wi...

Page 10: ...mages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibrat...

Page 11: ...it totale de FLIR est limit e la r paration ou au remplacement du produit La garantie d finie ci dessus est inclusive et aucune autre garantie crite ou orale n est exprim e ou implicite Calibrage r pa...

Page 12: ...entes La responsabilidad total de FLIR est limitada a la reparaci n o reemplazo del producto La garant a precedente es inclusiva y no hay otra garant a ya sea escrita u oral expresa o impl cita Servic...

Reviews: