background image

 

Oyster10-en-US_v2.3   9/15 

3

pH Overview 

pH is a unit of measure (ranging from 0 to 14pH) indicating the degree of acidity or alkalinity of a 
solution. pH tests are the most commonly performed measurements in water quality testing, using 
the hydrogen ion activity of a solution to determine acidity or alkalinity. Solutions with a pH less than 
7 are considered acidic, solutions with a pH higher than 7 are known as bases, and solutions with a 
pH of exactly 7 are considered neutral. 
The pH scale is logarithmic, so, for example, if sample A is 1 pH less than Sample B, this means 
that Sample A is 10 times more acidic then Sample B. A difference of 1 pH represents a ten-fold 
difference in acidity. 

ORP/REDOX Overview 

ORP stands for 

O

xidation 

R

eduction 

P

otential and represents the oxidizing or reducing nature of a 

solution. The overall “reactive” tendency of a solution can be determined from ORP measurements. 
ORP testing is becoming more common in waste-water and water treatment measurement 
regimens. The unit of measurement for ORP is mV.

 

Calibration and Operation 

pH Calibration (ORP does not require calibration)

 

Before accurate measurements can be made, it is necessary to perform a 

3-point calibration

 of the 

meter with its electrode. Calibrate to a pH 7 buffer first, followed by a pH 4 (or 1.68) buffer and then 
a pH 10 (or pH 12.45) buffer. The meter automatically recognizes the buffers listed in the table 
below and calibrates itself to the buffer’s pH value. 

1.  Connect the pH electrode to the meter’s BNC connector input. 
2.  Open the top half of the meter to an appropriate viewing position. 
3.  Power the meter with the 

POWER/CAL

 button. 

4.  Place the electrode into a buffer solution. 
5.  Press and hold the 

CAL

 key until the CAL display icon appears. 

6.  During calibration, ‘CAL’ is shown on the display followed by ‘SA’

 

(saved). 

7.  When calibration is complete, the meter automatically displays ‘END’ and returns to the normal 

operation mode. Repeat steps 1-5 as necessary. Note that it is not necessary to turn the meter 
off and then back on again to calibrate each reference. 

Error Message Note

: The meter will display ‘E13’ if a calibration step fails. Try the calibration again 

if ‘E13’ is displayed. The most common reason for the error is inaccurate or stale buffer solutions. 

pH reading 

Range of measurement 

1.68 

0.00 to 1.68 

4.00 

1.68 to 6.50 

7.00 

6.50 to 7.50 

10.00 

7.50 to 12.45 

12.45 

12.45 to 14.00 

 

Electrode note: 

Since pH electrodes can vary in performance, this meter displays an electrode’s 

slope in %. For example, if two electrodes (one new and one moderately used) are calibrated in the 
same pH 4 solution they’ll both display pH 4 on the meter. This is because the meter will accept one 
pH value for a range of measurement values as shown in the table above (in this case, any reading 
between 1.68pH and 6.50pH will be accepted as 4pH calibration). The meter then displays the 
slope % for the electrode under test. If the slope % is below 75% or above 125%, the electrode 
must be replaced. A slope of 100% is ideal. 

www.ponpe.com

contact:

[email protected]

Summary of Contents for Oyster-10

Page 1: ...USER GUIDE pH mV Temperature Meter Model OYSTER 10 www ponpe com contact sale ponpe com...

Page 2: ...lude diagnostic displays MIN MAX readings audible alerts 5 point buffer recognition Auto Manual temperature compensation This instrument is shipped fully tested and calibrated and with proper use will...

Page 3: ...calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 Open the top half of the meter to an appropriate viewing position 3 Power the meter with the...

Page 4: ...s complete unplug the electrode and rinse it in distilled water 8 Place approx 1 ml of pH 4 solution in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required...

Page 5: ...e MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This procedure is for...

Page 6: ...er or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible 7 When calibrat...

Page 7: ...ure readings LCD viewing angle 90 o to 180 o Over range Indication Displays Power 9 volt battery or optional AC adapter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery ico...

Page 8: ...nd federal regulations with regard to battery disposal Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of the same type Troubles...

Page 9: ...be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Long term storage wi...

Page 10: ...mages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibrat...

Page 11: ...it totale de FLIR est limit e la r paration ou au remplacement du produit La garantie d finie ci dessus est inclusive et aucune autre garantie crite ou orale n est exprim e ou implicite Calibrage r pa...

Page 12: ...entes La responsabilidad total de FLIR est limitada a la reparaci n o reemplazo del producto La garant a precedente es inclusiva y no hay otra garant a ya sea escrita u oral expresa o impl cita Servic...

Reviews: