Acquisition
Acquisition Modes Normal, Peak Detect, and Average
Acquisition Rate,
typical
Up to 2000 waveforms per second per channel (Normal acquisition
mode, no measurement)
Single Sequence
Acquisition Mode
Acquisition Stop Time
Normal, Peak Detect
Upon single acquisition on all
channels simultaneously
Average
After N acquisitions on all
channels simultaneously, N can
be set to 4, 8, 16, 32, 64 or 128
Inputs
Input Coupling
DC, AC or GND
Input Impedance,
DC coupled
1M
Ω
±2% in parallel with 20pF ±3pF
Probe Attenuation
1X, 10X
Supported Probe
Attenuation
1X, 10X, 100X, 1000X
Overvoltage Category
Maximum Voltage
CAT I and CAT II
300V
RMS
(10×), Installation
Category
CAT III
150V
RMS
(1×)
Maximum Input
Voltage
Installation Category II: derate at 20dB/decade above 100kHz to 13V
peak AC at 3MHz* and above. For non-sinusoidal waveforms, peak
value must be less than 450V. Excursion above 300V should be of less
than 100ms duration. RMS signal level including all DC components
removed through AC coupling must be limited to 300V. If these values
are exceeded, damage to the oscilloscope may occur.