58 MS6000-EU-EN-V1.3 7/12
4.5.7
Fast Action Buttons
AUTO:
Automatically set the oscilloscope controls to generate a usable display of the input signals.
Refer to the following table for relative content.
RUN/STOP:
Continuously acquire waveforms or stop the acquisition.
4.5.8
AUTOSET
Autoset is one of the most useful modes of the digital oscilloscope. When the
AUTO
button is pressed,
the oscilloscope will identify the type of waveform (sine or square) and adjust controls according to the
input signal so that it can accurately display the waveform.
Functions
Settings made automatically
Acquire Mode
Adjusts to Normal or Peak Detect
Cursor Off
Display Format
Sets to YT
Display Type
Sets to Vectors for an FFT spectrum; otherwise, unchanged
Horizontal Position
Adjusted
TIME/DIV Adjusted
Trigger Coupling
Adjusts to DC, Noise Reject, LF Reject or HF Reject
Trigger Holdoff
Minimum
Trigger Level
Sets to 50%
Trigger Mode
Auto
Trigger Source
Adjusted
Trigger Slope
Adjusted
Trigger Type
Edge
Trigger Video Sync
Adjusted
Trigger Video Standard
Adjusted
Vertical Bandwidth
Full
Vertical Coupling
DC (if GND was chosen before); AC for the video signal; otherwise,
unchanged
Volts Adjusted
www.burntec.com