76 MS6000-EU-EN-V1.3 7/12
DC Gain Accuracy
±3% for Normal or Average acquisition mode, 5V/div
to 10mV/div
±4% for Normal or Average acquisition mode,
5mV/div to 2mV/div
Measurement Type: Average of
≥
16 waveforms with
vertical position at zero
Accuracy: ± (3% × r 0.1div + 1mV) when
10mV/div or greater is selected
DC Measurement Accuracy,
Average Acquisition Mode
Measurement Type: Average of
≥
16 waveforms with
vertical position not at zero
Accuracy: ± [3% × (r vertical position) + 1%
of vertical po 0.2div]
Add 2mV for settings from 2mV/div to 200mV/div;
add 50mV for settings from 200mV/div to 5V/div
Volts Measurement Repeatability,
Average Acquisition Mode
Delta volts between any two averages of
≥
16
waveforms acquired under same setup and ambient
conditions
Note: Bandwidth reduced to 6MHz when using a 1X probe.