ACCURACY

 

SELF

TEST

 

FUNCTION

 

The

 

MG310

 

features

 

a

 

10M

Ω 

accuracy

 

check

 

feature

 

for

 

the

 

Insulation

 

Resistance

 

mode.

 

To

 

verify

 

the

 

accuracy:

 

1.

 

Connect

 

a

 

wire

 

jumper

 

to

 

the

 

meter

 

between

 

the

 

LINE

 

(4)

 

and

 

V

 

(3)

 

terminals

 

as

 

shown

 

in

 

the

 

accompanying

 

illustration.

 

2.

 

Select

 

one

 

of

 

the

 

three

 

test

 

voltage

 

(250/500/100V)

 

switch

 

positions.

 

3.

 

Press

 

the

 

TEST

 

button

 

to

 

begin

 

the

 

test.

 

4.

 

When

 

testing

 

is

 

complete,

 

the

 

meter

 

will

 

display

 

10.0M

Ω 

 

0.1M

).

 

If

 

the

 

meter

 

does

 

not

 

show

 

an

 

accurate

 

reading

 

please

 

have

 

it

 

serviced.

 

 

 

 

 

LOW

 

RESISTANCE

 

CONTINUITY

 

TESTING

 

WARNING: 

To avoid electric shock, never measure continuity on circuits or wires that have 

voltage on them. 

1.

 

Set

 

the

 

function

 

switch

 

to

 

the

   

(continuity)

 

position.

 

2.

 

Insert

 

the

 

black

 

test

 

lead

 

banana

 

plug

 

into

 

the

 

GUARD

 

(2)

 

jack.

 

Insert

 

the

 

red

 

test

 

lead

 

banana

 

plug

 

into

 

the

 

EARTH

 

(1)

 

jack.

 

3.

 

Press

 

the

 

TEST

 

button.

 

The

 

button

 

lamp

 

will

 

light

 

indicating

 

that

 

testing

 

can

 

begin.

 

4.

 

Touch

 

the

 

test

 

probe

 

tips

 

to

 

the

 

circuit

 

or

 

wire

 

you

 

wish

 

to

 

check.

  

5.

 

If

 

the

 

resistance

 

is

 

less

 

than

 

approximately

 

20

Ω

,

 

the

 

audible

 

signal

 

will

 

sound

 

and

 

the

 

audio

 

icon

 

will

 

be

 

displayed.

 

If

 

the

 

circuit

 

is

 

open,

 

the

 

display

 

will

 

indicate

 

>

 

220

Ω

.

 

6.

 

Press

 

the

 

TEST

 

button

 

again

 

to

 

disengage

 

the

 

test

 

mode

 

(the

 

lamp

 

will

 

switch

 

off)

 

before

 

removing

 

the

 

test

 

leads

 

from

 

the

 

circuit

 

under

 

test

 

and

 

from

 

the

 

meter.

 

 
 
 
 

M

Summary of Contents for MG310

Page 1: ...USER MANUAL Insulation Resistance Tester With Continuity and AC Voltage modes Model MG310...

Page 2: ...ch com website for newer versions of the user manual Misuse of this device may cause personal injury and damage to the instrument and connected equipment This symbol on the instrument indicates that f...

Page 3: ...scharged and isolated from power circuits before taking resistance measurements If the test leads or adaptors need to be replaced due to damage replace with test leads or adaptors of the same model or...

Page 4: ...ting selection button 8 TEST button 9 Rotary function select switch Note Secure battery compartment located on rear of unit Control Buttons 1 PI DAR button Press to select PI polarization index and DA...

Page 5: ...ction switch to the OFF position when the meter is not in use 2 If OL appears in the display during a measurement the value exceeds the meter s range 3 If the batteries are weak replace them before pe...

Page 6: ...be indicated in the right most auxiliary display the symbol will flash and the caution buzzer will sound 6 Press to release the TEST button to stop the test The high voltage will switch off and the re...

Page 7: ...test DIELECTRIC ABSORPTION RATIO DAR TESTING Read and understand all operations and safety information in the Insulation Resistance IR measurement section above and in the Safety section before conti...

Page 8: ...G To avoid electric shock never measure continuity on circuits or wires that have voltage on them 1 Set the function switch to the continuity position 2 Insert the black test lead banana plug into the...

Page 9: ...not measure voltage higher than 750V AC Please use extreme caution to avoid high voltage electrical shock while operating this instrument Disconnect test leads from tested circuits and meter terminal...

Page 10: ...or weak batteries so they do not leak and damage the unit 6 IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME the batteries should be removed to prevent damage to the unit BATTERY INSTALLATION W...

Page 11: ...5 5G 0 00M 5 5G 0 00M 5 5G Open circuit 250V DC 10 500V DC 10 1000V DC 10 Rated current 250K 1 00mA 1 10mA 500K 1 00mA 1 10mA 1M 1 00mA 1 10mA Short circuit Current Approx 2mA Accuracy 0 00M 99 9M 3...

Page 12: ...nd Automatic stop Test button Red lamp on TEST button indicates that test is in progress Self Test 10M 1 Continuity Check Audible signal when resistance is 20 approx Battery Six 6 1 5V AA batteries Cu...

Reviews: