7                                    MG310-en-GB_V1.0   12/15 

 

POLARIZATION

 

INDEX

 

(PI)

 

TESTING

 

Read

 

and

 

understand

 

all

 

operations

 

and

 

safety

 

information

 

contained

 

in

 

the

 

Insulation

 

Resistance

 

(IR)

 

measurement

 

section

 

above

 

and

 

the

 

Safety

 

section

 

before

 

continuing.

 

1.

 

Connect

 

the

 

meter

 

to

 

the

 

device

 

under

 

test

 

as

 

shown

 

in

 

the

 

test

 

section

 

above.

 

2.

 

Select

 

the

 

desired

 

output

 

test

 

voltage

 

using

 

the

 

rotary

 

function

 

switch

 

(250/500/1000V).

 

3.

 

Use

 

the

 

PI/DAR

 

button

 

to

 

access

 

the

 

PI

 

test

 

mode.

 

4.

 

The

 

display

 

will

 

show

 

the

 

PI

 

icon

 

when

 

the

 

PI

 

mode

 

is

 

successfully

 

accessed.

 

5.

 

The

 

two

 

test

 

times

 

(Time1

 

and

 

Time2)

 

will

 

alternately

 

show

 

on

 

the

 

upper

 

right.

 

6.

 

The

 

first

 

test

 

will

 

run

 

for

 

1

 

minute

 

(Time1)

 

when

 

the

 

TEST

 

button

 

is

 

pressed.

 

If

 

the

 

test

 

fails

 

(

no

 

is

 

displayed)

 

the

 

second

 

test

 

will

 

not

 

automatically

 

run

 

and

 

testing

 

is

 

now

 

complete.

 

If

 

the

 

test

 

passes

 

(

YES

 

is

 

displayed)

 

the

 

Time2

 

test

 

will

 

run

 

for

 

10

 

minutes.

 

7.

 

The

 

test

 

result

 

will

 

display

 

after

 

the

 

10

minute

 

Time2

 

interval

 

8.

 

Press

 

the

 

TEST

 

button

 

to

 

stop

 

the

 

test

 

and

 

to

 

begin

 

discharging

 

the

 

device

 

under

 

test.

 

 

DIELECTRIC

 

ABSORPTION

 

RATIO

 

(DAR)

 

TESTING

 

Read

 

and

 

understand

 

all

 

operations

 

and

 

safety

 

information

 

in

 

the

 

Insulation

 

Resistance

 

(IR)

 

measurement

 

section

 

above

 

and

 

in

 

the

 

Safety

 

section

 

before

 

continuing.

 

1.

 

Connect

 

the

 

meter

 

to

 

the

 

device

 

under

 

test

 

as

 

shown

 

in

 

the

 

IR

 

test

 

section

 

above.

 

2.

 

Select

 

the

 

desired

 

output

 

test

 

voltage

 

using

 

the

 

rotary

 

function

 

switch

 

(250/500/1000V).

 

3.

 

Use

 

the

 

PI/DAR

 

button

 

to

 

access

 

the

 

DAR

 

test

 

mode.

 

4.

 

The

 

display

 

will

 

show

 

the

 

DAR

 

icon

 

when

 

the

 

DAR

 

mode

 

is

 

successfully

 

accessed.

 

5.

 

The

 

first

 

two

 

test

 

times

 

(Time1

 

and

 

Time2)

 

will

 

alternately

 

show

 

on

 

the

 

upper

 

right.

 

6.

 

The

 

first

 

test

 

will

 

run

 

for

 

30

 

seconds

 

(Time1)

 

when

 

the

 

TEST

 

button

 

is

 

pressed.

 

If

 

the

 

test

 

fails

 

(

no

 

is

 

displayed)

 

the

 

second

 

test

 

will

 

not

 

automatically

 

run

 

and

 

testing

 

is

 

now

 

complete.

 

If

 

the

 

test

 

passes

 

(

YES

 

is

 

displayed)

 

the

 

Time2

 

test

 

will

 

run

 

for

 

1

 

minute.

 

The

 

test

 

result

 

will

 

display

 

after

 

the

 

1

minute

 

Time2

 

interval

 

7.

 

To

 

access

 

the

 

second

 

set

 

of

 

Time1

 

and

 

Time2

 

tests

 

press

 

the

 

DAR

 

button

 

three

 

times

 

and

 

repeat

 

the

 

same

 

steps

 

above

 

in

 

this

 

section.

 

For

 

the

 

2

nd

 

set

 

of

 

tests

 

the

 

times

 

are

 

15

 

seconds

 

(Time1)

 

and

 

1

 

minute

 

(Time2).

 

8.

 

When

 

completed

 

press

 

the

 

TEST

 

button

 

to

 

stop

 

the

 

test

 

and

 

to

 

begin

 

discharging

 

Summary of Contents for MG310

Page 1: ...USER MANUAL Insulation Resistance Tester With Continuity and AC Voltage modes Model MG310...

Page 2: ...ch com website for newer versions of the user manual Misuse of this device may cause personal injury and damage to the instrument and connected equipment This symbol on the instrument indicates that f...

Page 3: ...scharged and isolated from power circuits before taking resistance measurements If the test leads or adaptors need to be replaced due to damage replace with test leads or adaptors of the same model or...

Page 4: ...ting selection button 8 TEST button 9 Rotary function select switch Note Secure battery compartment located on rear of unit Control Buttons 1 PI DAR button Press to select PI polarization index and DA...

Page 5: ...ction switch to the OFF position when the meter is not in use 2 If OL appears in the display during a measurement the value exceeds the meter s range 3 If the batteries are weak replace them before pe...

Page 6: ...be indicated in the right most auxiliary display the symbol will flash and the caution buzzer will sound 6 Press to release the TEST button to stop the test The high voltage will switch off and the re...

Page 7: ...test DIELECTRIC ABSORPTION RATIO DAR TESTING Read and understand all operations and safety information in the Insulation Resistance IR measurement section above and in the Safety section before conti...

Page 8: ...G To avoid electric shock never measure continuity on circuits or wires that have voltage on them 1 Set the function switch to the continuity position 2 Insert the black test lead banana plug into the...

Page 9: ...not measure voltage higher than 750V AC Please use extreme caution to avoid high voltage electrical shock while operating this instrument Disconnect test leads from tested circuits and meter terminal...

Page 10: ...or weak batteries so they do not leak and damage the unit 6 IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME the batteries should be removed to prevent damage to the unit BATTERY INSTALLATION W...

Page 11: ...5 5G 0 00M 5 5G 0 00M 5 5G Open circuit 250V DC 10 500V DC 10 1000V DC 10 Rated current 250K 1 00mA 1 10mA 500K 1 00mA 1 10mA 1M 1 00mA 1 10mA Short circuit Current Approx 2mA Accuracy 0 00M 99 9M 3...

Page 12: ...nd Automatic stop Test button Red lamp on TEST button indicates that test is in progress Self Test 10M 1 Continuity Check Audible signal when resistance is 20 approx Battery Six 6 1 5V AA batteries Cu...

Reviews: