background image

                      MODEL 45168CP 

 

Model 45168CP-mul_V1.3 

 

Introduction 

Thank you for selecting the Extech 45168CP. This instrument measures Wind Speed, 
Temperature, RH%, Dew Point, Wet Bulb, and Wind Chill. The built-in 360

o

 Compass 

also provides wind direction readings. Features include delta ∆T (Air Temp. minus 
Dew Point), MAX/AVG wind speed, Auto Power Off, and water resistant housing.

 

Operation 

Opening the Meter 

Swivel the meter out from its protective case to a maximum 180 degree angle. Use 
a 45 degree angle for tripod use (tripod mount on bottom of meter). Close the 
meter when not in use. 

 

Turning the meter ON and OFF 

 

Press the   button to turn the meter ON

 

 

Press and hold the   button to turn the meter OFF

 

 

Auto Power OFF (APO) turns the meter OFF after 5 minutes of inactivity. To 
disable APO: With power OFF, press and hold both buttons until ‘n’ appears.

 

 

If the meter does not switch ON, please check the battery.

 

Select the Mode of Operation and the Unit of Measure 

 

With the meter ON, use the MODE (M) button to step: Wind speed > MAX wind 
speed > 10 sec. AVG wind speed > compass > air temperature > wind chill (WCI) 
> relative humidity (%RH) > wet bulb (WBT) > dew point (DP) > ∆T. 

 

In the Air Temperature mode, press the UNIT button to select 

o

F or 

o

C. 

 

In the Wind Speed mode, use the UNIT button to select unit of measure. 

Wind Speed Measurement Notes 

Position the meter so that the airflow enters the meter vane from the rear of the 
meter. A tripod mount is located on the bottom of the meter for convenience. 

Compass and Wind Direction (head/tail/cross) Modes 

 

Switch the meter ON and select the Compass mode

 

 

Point meter in the direction of travel and read the heading on the LCD. 

 

Press and hold the UNIT button until ‘head-tail-cross’ appears at the bottom of 
the LCD; the reading will then flash 3 times. Release the UNIT button. 

 

Point the meter into the wind until there is a steady compass reading. Press and 
hold the UNIT button until the compass reading flashes 3 times. 

 

The head or tail wind value will display. Press UNIT to see the crosswind. 

 

Press the 

M

 button to return to the compass-only mode. 

 

Note the wind direction indicator (item 2, Meter Description section). 

Wind Direction Considerations 

The head-tail-crosswind is the relationship between travel direction and wind 
direction. When wind direction is fixed and travel direction changes, the wind 
resistance changes, for example the wind resistance for 

A

 is stronger than 

B

 (see 

diagram) and 

C

 is stronger than 

D

. When calculating head, tail, and crosswind, first 

measure the travel direction (compass).  
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Compass Calibration 

 

In compass mode, press and hold the two buttons to 
access the 30 sec. timer. 

 

Rotate the meter twice slowly in the direction shown 
in the diagram. Each turn should be 15 seconds in 
duration. 

 

After the two turns, the LCD will show ‘END’ to confirm the 
calibration. 

 

Calibrate the meter before each use and always after a battery change. 

 

Meter Description 

 

1

 

Vane impeller (impeller set screw on back) 

2

 

Wind direction indicator 

3

 

Temperature and Humidity sensors 

4

 

LCD Display 

5

 

ON/OFF, M (MODE) button  

6

 

UNIT button 

7

 

Battery compartment (back) 

8

 

Tripod mounting hole 

9

 

Storage case 

10

 

Lanyard holder 

 

Maintenance 

Battery Replacement 

Turn the meter off before opening the battery compartment. Using a coin, turn the 
battery compartment cover (back) CLOCKWISE to remove it. Once opened, observe 
the position of the battery, placing the new one in the same position. Secure the 
battery compartment before use. Please dispose of the battery responsibly. Remove 
battery if meter is to be stored for a month or longer. 

EU users are legally bound by the Battery Ordinance to return all used batteries to 
community collection points or wherever batteries / accumulators are sold. 
Disposal in household trash or refuse is prohibited. 

Disposal: 

Follow the valid legal stipulations in respect of the disposal of the device at 

the end of its lifecycle 

Impeller Replacement 

 

Remove the rear set screw, located to the left of the impeller assembly.  

 

Twist the impeller assembly counter-clockwise to the OPEN (O) position and 
remove it. 

 

Install impeller by inserting & twisting the new impeller assembly clockwise to 
the LOCK (L) position. 

 

Tighten the set screw. 

 

CFM Air Volume Measurements 

 

Measure the area of the duct using the diagrams below for rectangular or circular 
duct. If the duct measurements are made in inches, divide the inches by 144 to get 
the area in square feet. Insert the area value (in square feet) into the equations 
below. Note that the air velocity must also be inserted into the cubic equations. 

 

 

 

 

 

 

CFM (ft

3

/min) = Air Velocity (ft/min) x Area (ft

2

CMM (m

3

/min) = Air Velocity (m/sec) x Area (m

2

) x 60 

 

Specifications 

Measurement 

Range 

Resolution 

Accuracy (% of rdg) 

MPH (miles per hour) 

0.5~44.7 MPH 

0.1 MPH 

± (3% + 0.4 mph) 

KPH (kilometers per hour)  0.8 to 72.0 km/h 

0.1 km/h 

± (3% + 1.4 km/hr) 

KNT (naut. miles/hour) 

0.4~38.8 knots 

0.1 knots 

± (3% + 0.6 knots) 

MPS (meters per second)  1.1~20.0 m/s 

0.1 m/s 

± (3% + 0.2 m/s) 

FPM (feet per minute) 

60~3937 ft/min 

2 ft/min 

± (3% + 40 ft/min) 

BF (Beaufort force) 

1~8 BF 

1 BF 

± 1 

Temperature 

-15~50

o

C (5~122

o

F) 

0.1

o

F/C 

± 1.0

o

C (± 1.8

o

F) 

Relative Humidity 

0.1~99.9% RH 

0.1% RH 

± 3% (10~90%) 

Dew Point Temperature 

-20~50

o

C (-4~122

 o

F)  0.1

o

F/C 

Calculation 

Wet Bulb Temperature 

-5~50

o

C (23~122

 o

F) 

0.1

o

F/C 

Calculation 

Compass 

0~360

o

 

1

o

 

± 2

 o

 

Wind Chill 

-20~50

o

C (-4~122

 o

F)  0.1

o

F/C 

± 2% 

Display 

LCD with multifunction indicators 

Sensors 

Sapphire bearing, non-corrosive vane; Precision thermistor for 

temperature measurements; Capacitive RH sensor 

AVG Mode 

10 reading averaging for wind speed mode 

MAX Mode 

MAX recalls the highest wind speed reading 

Response time 

Air Temperature and RH: 60 seconds (typical) 

Water and Drop Resistant 

Water resistant housing to 1m (3’) / Drop tested to 2m (6’) 

Operating conditions 

-15 to 50

o

C (5 to 122

o

F) / < 80% RH 

Power supply 

Lithium battery (CR-2032 or equivalent) 

Dimensions / Weight 

140 x 45 x 25mm (5.5 x 1.8 x 1.0”) folded / 90g (3.2 oz.) 
Vane diameter: 24mm (1.0”) 

Safety Standard 

EN 61326-1 (2013) 

 

Copyright 

©

 2015 FLIR Systems, Inc. 

All rights reserved including the right of reproduction in whole or in part in any form 

www.extech.com 

 
 
 

EN 

Summary of Contents for 45168CP

Page 1: ...set screw on back 2 Wind direction indicator 3 Temperature and Humidity sensors 4 LCD Display 5 ON OFF M MODE button 6 UNIT button 7 Battery compartment back 8 Tripod mounting hole 9 Storage case 10 Lanyard holder Maintenance Battery Replacement Turn the meter off before opening the battery compartment Using a coin turn the battery compartment cover back CLOCKWISE to remove it Once opened observe ...

Page 2: ...ulsor de veleta tornillo de ajuste del impulsor atrás 2 Indicador de dirección del viento 3 Sensores de temperatura y humedad 4 Pantalla LCD 5 Botón ON OFF M MODE 6 Botón UNIT 7 Compartimiento de la batería atrás 8 Orificio de montaje del trípode 9 Estuche 10 Orificio para correa Mantenimiento Reemplazo de la batería Apague el medidor antes de abrir el compartimiento de la batería Con una moneda g...

Page 3: ...pour confirmer le calibrage Calibrez l appareil avant chaque utilisation et toujours après le remplacement des piles Description de l appareil 1 Hélice du moulinet vis de blocage de l hélice située au dos de l appareil 2 Indicateur de la direction du vent 3 Capteurs de température et d humidité 4 Écran LCD 5 Bouton ON OFF M MODE 6 Bouton UNIT 7 Compartiment à pile au dos de l appareil 8 Trou pour ...

Page 4: ... Kalibrierung Kalibrieren Sie das Messgerät vor jeder Nutzung und nach jedem Batteriewechsel Übersicht des Messgeräts 1 Flügelrad Stellschraube an der Rückseite 2 Windrichtung Anzeige 3 Temperatur und Feuchtigkeitssensoren 4 LCD Display 5 EIN AUS M MODUS 6 UNIT Taste 7 Batteriefach Rückseite 8 Tripod Befestigung 9 Aufbewahrungstasche 10 Öse für Trageriemen Wartung Batterie ersetzen Schalten Sie da...

Page 5: ...so e sempre dopo la sostituzione della batteria Descrizione dell apparecchio 1 Girante della banderuola vite di regolazione della girante sul retro 2 Indicatore di direzione del vento 3 Sensori Temperatura e Umidità 4 Display LCD 5 Tasto ON OFF M MODE 6 Pulsante UNIT 7 Vano batteria retro 8 Foro di montaggio treppiede 9 Custodia 10 Supporto cordoncino Manutenzione Sostituzione delle Batterie Spegn...

Page 6: ...o do Medidor 1 Impulsor da aleta parafuso de ajuste do impulsor na traseira 2 indicador da direção do vento 3 Sensores de Temperatura e Umidade 4 Display LCD 5 Botão ON OFF ligar desligar M MODO 6 Botão UNIT Unidade 7 Compartimento da bateria na traseira 8 Furo de montagem do tripé 9 Caixa de armazenamento 10 Suporte do colhedor Manutenção Substituir a Bateria Desligue o aparelho antes de abrir o ...

Reviews: