380366  V1.6  11/04

 

Operation 

 

 

 

 

 

Connecting Test Leads 

For all functions (Megohmmeter, Resistance, Continuity and AC Voltage), connect the red 
test lead to the HI terminal and the black lead to the LO terminal. 

Test Lead Check 

1.  Connect the red test lead to the HI input terminal; black lead to the LO terminal. 

2.  Set the rotary switch to the resistance/ continuity position ( .    

Ω)

 

3.  Touch the test lead tips together. Resistance should be less than 1

 and the audio 

tone should sound. 

4.  With the leads not touching, the display should read “-OL-“.  (over range). 
5.  If any readings are displayed other than the readings just described, the test leads 

should be considered faulty and must be replaced before using the meter. Failure to 
do so could result in damage to equipment and electrical shock. 

Insulation Resistance Measurements (Megohmmeter Tests) 

 

 

1.  Connect the red test lead to the HI input terminal; black lead to the LO terminal. 
2.   Set the rotary switch to the desired function. 
3.  Connect the tips of the test leads to the equipment under test. 
4.  Press and hold down the red TEST button. Release the test button to stop the test. The 

meter will then hold the resistance reading. This is indicated by the H icon in the upper 
left corner of the display. 

5.  To clear the display, press the LOCK/HOLD button twice. 
 Note that the charge stored on the device under test will be automatically discharged  
when the testing process is finished 

3-Minute Test 

For hands-free operation, use the Test Lock feature. 
1. 

Connect the red test lead to the HI input terminal; black lead to the LO terminal  

2. 

Set the rotary switch to the desired function. 

3. 

Connect the tips of the test leads to the equipment under test. 

4. 

Press the LOCK button. The LOCK icon will be displayed. Press the red TEST button. 
A 3-minute continuous test will begin. The red TEST button can be pressed at any 
time to end the test. If the TEST button is not pressed the meter automatically stops 
the test after 3 minutes. The meter will then hold the resistance reading. This is 
indicated by the H icon in the upper left corner of the display. 

5. 

To clear the display, press the LOCK/HOLD button twice. 

 

Note:

 The         symbol will appear in the upper right corner of the display and an audible 

tone will sound whenever the high voltage is applied to the test leads. 

Warning 

Ensure that the circuit under test does not include components that can be damaged by 1000VDC; such 

devices include power factor correction capacitors, low voltage mineral insulated cables, electronic light 
dimmers, and ballasts/starters for fluorescent lamps. 

Warning: Do not perform Insulation Resistance measurements if AC Voltage is present on circuit under test. 

Summary of Contents for 380366

Page 1: ...MΩ and includes Continuity and Voltage tests The built in datalogger remotely stores approx 16 000 readings for later transfer to a PC The tester can also take and record readings to a PC in real time This professional tester with proper care will provide years of safe reliable service 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDe...

Page 2: ...inforced insulation Specifications General specifications Display Backlit 3 3 4 digit 4000 count LCD with analog indicator Sampling rate 2 5 samples sec digital 10 samples sec analog Audible Continuity Low resistance range with audible continuity 40Ω Over range indicator OL displayed when measurement exceeds range limits Low battery indicator symbol displayed when battery voltage is low Datalogger...

Page 3: ...Voltage Ranges Maximum Resolution Resistance Accuracy Voltage Accuracy 250V 20 0 500V 20 0 1000V 4 40 400 4000MΩ 1kΩ 3 rdg 5d 2000MΩ 5 rdg 5d 2000MΩ 10 0 Voltage Range 1mA Test Current Short Circuit Current 250V 250KΩ 500V 500KΩ 1000V 4000MΩ 1MΩ 1 5mA Note Accuracies are specified as reading digits at 73 F 9 F 23 o C 5 o C 80 RH Meter Description 1 LCD Display 2 Megohmmeter TEST pushbutton 3 Test ...

Page 4: ...play 5 To clear the display press the LOCK HOLD button twice Note that the charge stored on the device under test will be automatically discharged when the testing process is finished 3 Minute Test For hands free operation use the Test Lock feature 1 Connect the red test lead to the HI input terminal black lead to the LO terminal 2 Set the rotary switch to the desired function 3 Connect the tips o...

Page 5: ...esistance Measurements Megohmmeter tests but for the other tests the user must press the HOLD key to freeze the meter reading Press HOLD again to exit the Data Hold mode Auto Power OFF Feature After a 30 minute period of non use the meter automatically shuts down This will conserve battery life To wake the meter press the Backlight button Note that the meter is not completely off until the functio...

Page 6: ...apturing one data point Press the RECORD button for one second to capture one reading The REC display icon will briefly appear at the top of the LCD and the large digits will flash the data point number the recorded measurements are numbered Each reading captured by this method is considered a Set c Capturing a continuous stream of data points 1 Press and hold the RECORD button for approx 3 second...

Page 7: ...e program icon 2 Select PC communications port COM1 or COM2 when prompted 3 The Program screen will appear as a facsimile of the meter s front panel see below Data Acquisition and Datalogger modes These two modes of operation distinctly different are explained in detail separately For the Data Acquisition mode of operation the meter must be connected to the PC In this mode readings taken by the me...

Page 8: ...re how often to log a reading The sampling time can be set from 1 to 65 400 seconds See the LOGGER screen at right 1 Click on the wrist watch icon from the row of Control Icons on the main software screen 2 Highlight the existing value in the SAMPLING field 3 Type in the desired sampling time 4 Click the Sampling button when finished Creating a File to store data 1 Click the floppy disk icon from ...

Page 9: ...l Use the four front panel keys on the main software screen to remotely control the meter TEST button Click to start a megohmmeter test LOCK button Click to lock the test for 3 minutes read about the LOCK function earlier in this manual HOLD button Click to freeze the displayed reading The H display icon will appear while the meter is in the Data Hold mode Press HOLD again to exit this mode REC bu...

Page 10: ...download all of the data sets from the meter to the PC The data lists will appear as shown on the logger window diagram below 4 Select a set of data by clicking on the set number The NUMBER OF SET field informs the user as to which set is selected The NUMBER OF RECORDS field lets the user know the number of readings in the selected set 5 The MEMORY and REMAINING fields inform the user as to the to...

Page 11: ... on the commutator 3 If the resistance measurement indicates a weakness raise the brushes off of the commutator and separately test the armature field coils and brush rigging one at a time Leave one lead connected to the grounded motor housing while testing the motor components This also applies to DC Generators Testing Cables 1 Disconnect the cable under test from the line 2 Disconnect the opposi...

Reviews: