Using the XTA-50 in Remote Mode
54
XTA-50 User Guide
5.8.2.3
Remote Mode Control
LOCAL
Syntax
LOCAL
Description
Ends remote mode, the XTA-50 comes back to local mode.
ETH_IP=
Syntax
ETH_IP=<value>.<value>.<value>.<value>
Description
Sets a new IP address (for more details, see section
Setting the XTA-50 Ethernet Ports,
). The new value is only taken into account after calling
ETH_MANUAL
.
We do not recommend to set this value using the Ethernet remote mode.
•
<value>
: integer in the range 0 to 255.
Example
ETH_IP=192.168.0.5
ETH_IP?
Syntax
ETH_IP?
Description
Queries the XTA-50 IP address.
Query Response
String of four integers (0 to 255) separated by dots
Example
ETH_IP?
→
192.168.0.5
ETH_MASK=
Syntax
ETH_MASK=<value>.<value>.<value>.<value>
Description
Set a new mask for Ethernet connection (for more details, see section
). The new value is only taken into account after calling
ETH_MANUAL
.
We do not recommend to set this value using the Ethernet remote mode.
•
<value>
: integer in the range 0 to 255.
Example
ETH_MASK=255.255.255.0