Description of Functions
CT440
113
Configuration Functions
CT440_MeasureDark
Applicability
All models of CT440.
Description
This function measures the power on the selected detector and
stores it for power correction.
The measured power is automatically taken into account for
correction in the next power measurement.
In the GUI, this function is identical to the
Zero Detector X
buttons in
the
Referencing
window.
Declaration
int32_t CT440_MeasureDark (uint64_t uiHandle,
rDetector eDetector);
Parameters
uiHandle
Handle returned from
CT440_Init
Type: Input
Data type: unsigned int64
eDetector
Front panel detector number.
Possible values: DE_1; DE_2; DE_3; DE_4.
Type: Input
Data type: rDetector
Return value
0: the operation succeeded.
-1: the operation failed because of invalid uiHandle
-2: the operation failed because of missing calibration file
-3: the operation failed because of too much power on is
measured on the detector
Data type: int32
Example
CT440_MeasureDark (uiHandle,DE_1);
Summary of Contents for CT440 Series