Display
Message
Definition
Description
XXXX.
Lamp Hours
Displays the “hours of use” accumulated by the
lamp
Flashing:
“
XXXX
”
Warming Up
The lamp is warming up. Warm up time is
approximately 4 minutes ref
section 6.1
bulb
Lamp Error
Lamp installed incorrectly /Lamp did not strike.
See
section 8.3
cool
Lamp is too hot to strike
The lamp will automatically strike when it has
cooled to the optimum striking temperature.
See
section 8.3
Alternating:
old / bulb
Old Lamp
The lamp has accumulated over 2500 hours.
Lamp may be near end of life.
Alternating:
end / bulb
New Lamp Required
The lamp has reached end of life. The lamp
will not strike.
SFI
Shutter Failure
The shutter has failed to return to home
position. Unit should be restarted. If the error
repeats,
E1
Lamp temperature is too high
Check ventilation filters and outlets to ensure
that no blockage exists. Restart unit and if
problem persists replace lamp. If problem
continues
E4
Internal hardware failure
Restart unit, if problem persists
E5
Failure to communicate with
Intelli-Lamp
Check Intelli-Lamp connection and restart unit.
If problem persists try another lamp, if problem
still persist
E6
Lamp self extinguished
Let system cool and restart lamp. If it goes out
again, replace lamp. If new lamp continues to
self-extinguish,
E7
Internal system error
Restart unit, if problem persists
See Section 8 for troubleshooting.