![Excelitas Technologies 010-00369R Installation Reference Manual Download Page 33](http://html1.mh-extra.com/html/excelitas-technologies/010-00369r/010-00369r_installation-reference-manual_2449424033.webp)
OmniCure® LX500 Series
UV Curing Solutions
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2020
All rights reserved
035-xxxxxR, rev.1
29
035-00628R
Item
Description
3 – Foot Pedal 3
4 – Foot Pedal 4
5 – PLC Input 1
6 – PLC Input 2
7 – PLC Input 3
8 – PLC Input 4
9 – PC
Repeat
The number of times the program is to be repeated from 2>10.
Calibration
The calibration type the intensity value is for. If 0, the intensity value is % setting. If 1 the
intensity value is irradiance in W/cm2. If 2, the intensity value is power in W.
Log File
The log file is auto generated; the title row has the description for the columns as well as the following table:
Item
Description
Date
Date the log entry was added.
Time
Time the log entry was added.
Description
An English text description of the event.
Hd 1 Status
Error code for the head, see the error code description table for numbers to description.
Hd 1 Time
The exposure time setting in seconds.
Hd 1 Pwr
If a decimal number, the irradiance in W/cm2, if a whole number, the percentage setting.
Hd 1 Cnt
The exposure time count in seconds, that the head was in when the event occurred. E.g.
if 5.0, it had 5.0 seconds remaining on the exposure.
Hd 1 Temp
The temperature of the head when the event occurred.
Hd 2 Status
Error code for the head, see the error code description table for numbers to description.
Hd 2 Time
The exposure time setting in seconds.
Hd 2 Pwr
If a decimal number, the irradiance in W/cm2, if a whole number, the percentage setting.
Hd 2 Cnt
The exposure time count in seconds that the head was in when the event occurred. E.g. if
5.0, it had 5.0 seconds remaining on the exposure.
Hd 2 Temp
The temperature of the head when the event occurred.
Hd 3 Status
Error code for the head, see the error code description table for numbers to description.
Hd 3 Time
The exposure time setting in seconds.
Hd 3 Pwr
If a decimal number, the irradiance in W/cm2, if a whole number, the percentage setting.
Hd 3 Cnt
The exposure time count in seconds that the head was in when the event occurred. E.g. if
5.0, it had 5.0 seconds remaining on the exposure.
Hd 3 Temp
The temperature of the head when the event occurred.
Hd 4 Status
Error code for the head, see the error code description table for numbers to description.
Hd 4 Time
The exposure time setting in seconds.
Hd 4 Pwr
If a decimal number, the irradiance in W/cm2, if a whole number, the percentage setting.
Hd 4 Cnt
The exposure time count in seconds, that the head was in when the event occurred. E.g.
if 5.0, it had 5.0 seconds remaining on the exposure.
Hd 4 Temp
The temperature of the head when the event occurred.
System Status
The error code for the system, see the error code descriptions for number to description.
Calibration Log File
The calibration log file keeps a record of all calibration performed on the system. Each time a calibration is
performed a new record is added to the log file.