OmniCure® LX500 Series
UV Curing Solutions
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2020
All rights reserved
035-xxxxxR, rev.1
23
035-00628R
7.2
Error Codes
Error
Description
Solution
Bit 1
(1)
The head(X) has exceeded the allowable operating
temperature. (error code will remain present until cleared,
however the head will be operative again when the
temperature is below the maximum allowable). The channel
select button is not required to clear this error. Subsequent
triggers will clear the alarm if the temperature of the LED has
recovered to below the over-temperature cut-off threshold.
(This applies to the front panel Start/Stop button, Foot Pedal,
and PLC I/O Start/Stop inputs).
Check position of clamp
location on the head. Refer to
Figure 5.
Check emission mode setting.
Check timing mode setting.
Check lens cleanliness.
Bit 2
(2)
The head(X) has a short circuit.
Check LED head to controller
connections.
Bit 3
(4)
The head (X) has been disconnected after power on. (The
head will remain disabled until the power is cycled).
Check LED head to controller
connections.
Bit 4
(8)
The head is open circuit. (The head will remain disabled until
the power is cycled).
Check LED head to controller
connections.
Check LED head wire for
damage.
Bit 5
(16)
The 1-wire memory chip for head (X) has an error. (The head
will remain disabled until the power is cycled).
Check LED head to controller
connections.
Bit 6
(32)
No temperature reading from head (X). (The head will be
deactivated until the power is cycled).
Check LED head to controller
connections.
Bit 7
(64)
Head is in a different controller serial # or different head
connector position.
Power off unit and plug into
controller serial # and head
position specified by error
message.
Press select to clear this error
and accept the new
connection.
Bit 8
(128)
Intensity setting in the StepCure profile exceeds the
capability of the head. Only occurs when the head is
calibrated, and the StepCure setting is an irradiance or
power setting. This also can occur if the setting in StepCure
is an irradiance setting, and the head is calibrated in power,
or vice versa.
Adjust the intensity setting in
the StepCure profile.
Recalibrate the head, ensure
that the peak intensity is found
prior to calibration.
Err1
Internal hardware error
Reset system.
If error code persists contact
service center.
Err2
Internal Communication failure
Reset system.
If error code persists contact
service center.
Err3
Non-Volatile Memory error
Reset system.
If error code persists contact
service center.
Err4
Not used
Not used
Err5
Inadequate system performance
The system cannot keep up
with the control inputs, slow
down the control inputs for
proper operation.