8.3
Cleaning the LED Head and Lens Assembly
**WARNING**
Ensure power is OFF and the UV LED Head is cool before attempting any cleaning
procedure.
Regular cleaning of the optical surface is recommended for optimal performance.
Wear powder free latex gloves or finger cots suitable for the cleaning solution selected to protect hands from
solvents and prevent window from contamination. Refer to the MSDS for guidance.
Blow off dirt and dust with pressurized air on front glass surface.
Use a lens tissue folded into quarters or a lint-free cotton swab. Saturate the lens tissue or swab with organic
solvent as indicated in
Section 8.2
,
“Cleaning Materials”
.
Using a new saturated tissue; clean LED Head by gently wiping the surface with a continuous stroke in one
direction.
Note:
Never apply a dirty tissue or swab to the UV LED Head.
Note:
Never face the LED head upwards before the solvent has evaporated.
8.4
Cleaning the Controller
**CAUTION**
Do not use any alcohol- or acetone-based solutions to cleaning the controller
housing.
Ensure the controller is in the off position and the power supply is unplugged.
Use a mild soap solution as the cleaning media.
Dampen a clean lint free cloth, tissue or swab with the mild soap solution and wipe the exterior of the housing.
Avoid excessive moisture when wiping in the air vent areas.
Using a new dampened tissue (or pre-saturated tissue), clean the optical window by gently wiping the surface
with one continuous stroke in one direction.
Note:
Never apply a dirty tissue or swab to the optical window of the controller.
Inspect the window to verify no residue is present.
8.5
Calibrating a UV LED Head
**CAUTION**
Do not stare directly at the light emitted from the LED array. This may be harmful,
resulting in eye injury. Always use UV protective eyewear.
The calibration process requires an Excelitas Technologies LS100 radiometer, or an LS100P radiometer. The LS100
is an irradiance only sensor that will calibrate the heads in W/cm2. The LS100P is a power only sensor that will
calibrate the heads in W. Ensure the usage of the proper sensor for your application.
To calibrate the UV LED Heads connected to the LX500 controller:
1.
Securely fixture the UV LED head you wish to calibrate.