ERROR & REPAIR
Error code
Cause
Repair
38001
• Electrode and measuring parts are not detected or dirty
• Faulty in impedance measuring
• Try to measure again after cleaning the electrode holders with soft gauze.
• Try again after cleaning the hands and soles.
• Check if foreign materials are between electrodes and measuring parts.
• Try again in correct posture and hold the electrodes according to the measuring
method.
38003
• There is an error in measured impedance by
external influence.
• Check installation condition (AC cord or around equipment
• Contact our company or distributor.
38004
• Out of measurement range of impedance
• range: 100~950Ω
• Try again after cleaning the hands and soles.
• Check if foreign materials are between electrodes and measuring parts.
• It can’t measure if the impedance is out of range.
38005
• Out of measurement range of PBF
• range: Under 75%
• Try again after cleaning the hands and soles.
• It can’t measure if the PBF is out of range.
38006
• Continuous error for 3 times
• Step off and try again.
• Contact our company or distributor.
38007
•
Disconnect Color board and Main board
• Contact our company or distributor.