PROHIBITION
PROHIBITION
Do not disassemble or alter the device under any circumstances, as this could
result in electric shock or injury as well as adversely affect the precision of
measurements.
This device is specified as Class 1 type BF unit under the standard IEC 60601-1:
2005+A1:2012 (Basic safety and essential performance of Medical Electrical
Equipment). Therefore, patients must not touch or handle inner side of the
system at any time.
PROHIBITION
PROHIBITION
Do not to touch signal input, signal output or other connectors, and the patient simultaneously.
The unit has previously been adjusted in the factory for opkmum performance.
Do not adempt to adjust switches or any other things except those specified in this manual for operakon.
Never pour any liquid directly on the scale plamorm, as it may leak and cause internal damage.
Never jump on the Weighing Platform, there may be a risk of stumbling and malfunction of
the equipment.
PROHIBITION
NOTE
This equipment has been tested and found to comply with the limits for medical devices according to
IEC 60601-1-2:2014. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation. This equipment
generates uses and can radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to other devices in the vicinity. However, there
is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to other devices, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are
connected.
- Consult the manufacturer or field service technician for help.
NOTE
NOTE
NOTE
Place the Weighing Plamorm on a level and stable surface. If the equipment is used when the
Weighing Plamorm is unstable because not all feet are on the
surface, there may be a risk of stumbling or inaccurate measurement.
Note that portable and mobile RF communications equipment can affect
MEDICAL ELECTRICAL EQUIPMENT.
Consult a physician or a trained health professional for interpretakon of measurement results.