39
38
Performing a Quality Control Test
9. Tightly replace the cap on the
Control Solution GC700 bottle.
10. The control solution result will appear. Compare your Quality Control Test result
to the Control Solution Range printed on the
Blood Glucose Test Strip
GS700 vial label.
Performing a Quality Control Test
CLOSE
9
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however,
they can still can be recalled and viewed. The Control Solution Test result will be shown
with the " " icon on the screen.
- The Control Solution Test should be conducted between
(
).
- Before " " and " " appear, do not touch the control solution to the sample port on the
strip. The
Blood Glucose Meter GM700S is performing an internal check.
Touching the control solution to the sample port before prompted will result in an error
message: " " and " "
.
- Do not drip the control solution on to the sample port of the test strip directly. The reagent
on the strip might leak into the bottle of control solution and may cause the degeneration
of the control solution. This could contaminate the meter via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is
stained or is overly exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean
with water.
6-44 °C 43-111 °F
and be accompanied by beeps (if volume is turned on)
1. Take one
Blood Glucose Test Strip GS700 from the vial and
close the vial cap immediately.
2. Insert the strip into the test strip port of the
Blood Glucose
Meter GM700S with the view window facing up.
3. While the blood drop icon is flashing on the display window, press
and hold the Main button for at least 3 seconds until the " "
symbol appears.
4. You will see the blinking "
" icon and " " icon on the screen,
prompting you to apply the corresponding level of
Control
Solution GC700.
5. Shake the bottle of
Control Solution GC700 well before
opening the cap. Place the cap on a flat surface.
6. Place a drop of control solution onto the top of the cap.
7. Gently touch the sample port of the strip with the control solution from
the top of the cap.
8.
The screen will display the count time starting from 5 (you will hear a
beep if the volume is turned on).
1
2
8
OPEN
5
6
3
GM700S
GM700S
7
GM700S
Summary of Contents for GM700S
Page 2: ......